Systematic evaluation of velocity-selective arterial spin labeling settings for placental perfusion measurement
Anita A Harteveld1, Jana Hutter2, Suzanne L Franklin1,3
1Department of Radiology, University Medical Center Utrecht, Utrecht University, Utrecht, the Netherlands.
Purpose:
Placental function is key for successful human pregnancies. Perfusion may be a sensitive marker for the in vivo assessment of placental function. Arterial spin labeling (ASL) MRI enables noninvasive measurement of tissue perfusion and it was recently suggested that ASL with velocity-selective (VS) labeling could be advantageous in the placenta. We systematically evaluated essential VS-ASL sequence parameters to determine optimal settings for efficient placental perfusion measurements.
Methods:
Eleven pregnant women were scanned at 3T using VS-ASL with 2D multislice echo planar imaging (EPI)-readout. One reference VS-ASL scan was acquired in all subjects; within subgroups the following parameters were systematically varied: cutoff velocity, velocity encoding direction, and inflow time. Visual evaluation and region of interest analyses were performed to compare perfusion signal differences between acquisitions.
Results:
In all subjects, a perfusion pattern with clear hyperintense focal regions was observed. Perfusion signal decreased with inflow time and cutoff velocity. Subject-specific dependence on velocity encoding direction was observed. High temporal signal-to-noise ratios with high contrast on the perfusion images between the hyperintense regions and placental tissue were seen at ~1.6 cm/s cutoff velocity and ~1000 ms inflow time. Evaluation of measurements at multiple inflow times revealed differences in blood flow dynamics between placental regions.
Conclusion:
Placental perfusion measurements are feasible at 3T using VS-ASL with 2D multislice EPI-readout. A clear dependence of perfusion signal on VS labeling parameters and inflow time was demonstrated. Whereas multiple parameter combinations may advance the interpretation of placental circulation dynamics, this study provides a basis to select an effective set of parameters for the observation of placenta perfusion natural history and its potential pathological changes.