Related Experiment Video
Updated: Oct 13, 2026

Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
Screening for arrhythmia with the new portable single-lead electrocardiographic device (SnapECG): an application
Jieqiong Guan1, Ailian Wang2, Wenjing Song1
1Key Laboratory of Environmental Medicine Engineering, Ministry of Education, Department of Epidemiology and Biostatistics, School of Public Health, Southeast University, 87 Ding Jiaqiao Rd., Nanjing, 210009, China.
Background:
SnapECG is a new handheld single-lead electrocardiograph (ECG) device used for arrhythmia screening, it is widely used in clinical practice but not in primary care.
Aims:
To evaluate the arrhythmia screening value of SnapECG among a community-based population.
Methods:
A cross-sectional community-based study of multistage stratified cluster sampling was conducted from March 1st to April 30th 2019. The sensitivities, specificities and the area under the receiver operating characteristic (AUCROC) curves of the SnapECG and reference 12-lead ECG on arrhythmia were calculated in three age-groups [50-64 years, 65-74 years, and over-75 years].
Results:
A total of 2263 participants took part in the arrhythmia screening, these included 1479 aged 50-64 years, 602 aged 65-74 years and 182 aged over-75 years. The SnapECG categorized 1828 (80.8%) as sinus rhythm, 161 (7.1%) as premature atrial/ventricular contractions (PAVs/PCVs), 32 (1.4%) as possible atrial fibrillation (AF), 56 (2.5%) as supraventricular tachycardias or sinus bradycardia (SVT/SB) and 186 (8.2%) as unreadable. SnapECG had 89% sensitivity (95% CI 0.52-1.00) and 99% specificity (95% CI 0.97-0.99) of detecting AF in the 65-74 years age-group. The AUCROC to detect AF was 0.94 for the 65-74 years age-group, 0.77 for over-75 years, 0.62 for the 50-64 years.
Discussion:
This study is the first community screening application of SnapECG. Main limitation is the SnapECG and the 12-lead ECG were not done simultaneously.
Conclusions:
In the people aged 65-74 years, AF can be detected accurately by the SnapECG with high sensitivity, specificity and large area under the ROC curve, which might have the highest screening predictive accuracy.
More Related Videos
08:10Estimating Bilateral Atrial Function by Cardiovascular Magnetic Resonance Feature Tracking in Patients with Paroxysmal Atrial Fibrillation
Published on: July 20, 2022
07:12Studying the Coding Profiles of Somatic Stimulation on Cardiac-locked Neuronal Responses in the Rat Spinal Dorsal Horn
Published on: May 23, 2025
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and the T...
Holter Monitor: 24-Hour Monitoring
Dysrhythmias V: Evaluating Dysrhythmias