Active control of selectivity in organic acid analysis by gas chromatography
Ernest Darko1, Kevin B Thurbide1
1Department of Chemistry, University of Calgary, 2500 University Drive, NW, Calgary, Alberta, T2N 1N4, Canada.
Abstract:
A new method that allows organic acid selectivity to be dynamically controlled during gas chromatography (GC) is presented. It employs dual in-series stainless steel columns, each coated with a pH-adjusted water stationary phase. The first is a 2 m column coated with a pH 11.4 phase that is connected to a second 11 m column coated with a pH 2.2 phase. In this arrangement, organic acids within sample mixtures are trapped on the first column, while the remaining non-ionizable components continue to separate and elute in the system. Later, by injecting a volatile formic acid solution, the trapped acids are released in-situ to the second column for separation and analysis as desired. The method provides good reproducibility with analyte retention times in consecutive trials yielding an average RSD of 1.9%. Further, depending on column temperature, analytes can be readily retained for periods investigated up to about 30 min without significant deterioration in peak shape. This feature provides considerable control over analyte selectivity and resolution compared to conventional separations. Further, by adding a third conventional GC column in-series, both typical hydrocarbon and enhanced organic acid separations are made possible. The method is applied to the analysis of complex mixtures and matrix interference is found to be significantly minimized. Results indicate that this approach offers beneficial advantages for the selective GC analysis of such acidic analytes.
More Related Videos
08:15Temperature-programmed Deoxygenation of Acetic Acid on Molybdenum Carbide Catalysts
Published on: February 7, 2017
08:37Measurement of H2S in Crude Oil and Crude Oil Headspace Using Multidimensional Gas Chromatography, Deans Switching and Sulfur-selective Detection
Published on: December 10, 2015
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...