Validation of an MR-guided online adaptive radiotherapy (MRgoART) program: Deformation accuracy in a heterogeneous,
Kathryn E Mittauer1, Patrick M Hill2, Michael F Bassetti2
1Miami Cancer Institute, Baptist Health South Florida, Miami, United States; Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin-Madison, Madison, United States.
Background And Purpose:
To investigate deformable image registration (DIR) and multi-fractional dose accumulation accuracy of a clinical MR-guided online adaptive radiotherapy (MRgoART) program, utilizing clinically-based magnitudes of abdominal deformation vector fields (DVFs).
Materials And Methods:
A heterogeneous anthropomorphic multi-modality abdominal deformable phantom was comprised of MR and CT anatomically-relevant materials. Thermoluminescent dosimeters (TLDs) were affixed within regions of interest (ROIs). CT and MR simulation scans were acquired. CT was deformed to MR for dose calculations. MRgoART was executed on a MR-linac (MRIdian) for 5 Gy/5 fractions. Before each fraction, a deformation was applied. Ground truth was known for ROI volume, TLD position, and TLD dose measured by an accredited dosimetry calibration laboratory. To validate the range of applied deformations, phantom DVFs were compared to DVFs of clinical abdominal MRgoART fractions. MR-MR deformation accuracy was quantified through dice similarity coefficient (DSC), Hausdorff distance (HD), mean distance-to-agreement (MDA), and as mean-absolute-error (MAE) for CT-MR-MR deformation. Arithmetic-summation of calculated dose at respective TLD positions and deform-accumulated dose (MIM) was compared to TLD measured dose, respectively. MR-MR deformation statistics were quantified for MRIdian and MIM.
Results:
Mean phantom DVFs were 5.0 ± 2.9 mm compared to mean DVF of clinical abdominal patients at 5.2 ± 3.0 mm. Respective mean DSC, HD, MDA was 0.93 ± 0.03, 0.74 ± 0.80 cm, 0.08 ± 0.03 cm for MRIdian and 0.93 ± 0.03, 0.54 ± 0.27 cm, 0.08 ± 0.03 cm for MIM (N = 80 ROIs). Mean MAE was 20.5 HU. Respective mean and median dose differences were 0.3%, -0.3% for arithmetic-summation and 4.1%, 0.6% for deformed-accumulation. Maximum differences were 0.21 Gy (arithmetic-summation), 0.31 Gy (deformed-accumulation).
Conclusions:
MRgoART deformation and dosimetric accuracy has been benchmarked for mean fractional DVFs of 5 mm in a multiple-rigid-body deformable phantom. Deformation accuracy was within TG132 criteria and clinically acceptable end-to-end MRgoART dosimetric agreement was observed for this phantom. Further efforts are needed in validation of deform-accumulated dose.