Related Experiment Video
Updated: Dec 26, 2025

Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Residual volatile anesthetics after workstation preparation and activated charcoal filtration
Lukas M Müller-Wirtz1, Christine Godsch1, Daniel I Sessler2
1Center of Breath Research, Department of Anesthesiology, Intensive Care and Pain Therapy, Saarland University Medical Center, Homburg, Saarland, Germany.
Background:
Volatile anesthetics potentially trigger malignant hyperthermia crises in susceptible patients. We therefore aimed to identify preparation procedures for the Draeger Primus that minimize residual concentrations of desflurane and sevoflurane with and without activated charcoal filtration.
Methods:
A Draeger Primus test workstation was primed with 7% desflurane or 2.5% sevoflurane for 2 hours. Residual anesthetic concentrations were evaluated with five preparation procedures, three fresh gas flow rates, and three distinct applications of activated charcoal filters. Finally, non-exchangeable and autoclaved parts of the workstation were tested for residual emission of volatile anesthetics. Concentrations were measured by multicapillary column-ion mobility spectrometry with limits of detection/quantification being <1 part per billion (ppb) for desflurane and <2.5 ppb for sevoflurane.
Results:
The best preparation procedure included a flushing period of 10 minutes between removal and replacement of all parts of the ventilator circuit which immediately produced residual concentrations <5 ppm. A fresh gas flow of 10 L/minute reduced residual concentration as effectively as 18 L/minute, whereas flows of 1 or 5 L/minute slowed washout. Use of activated charcoal filters immediately reduced and maintained residual concentrations <5 ppm for up to 24 hours irrespective of previous workstation preparation. The fresh gas hose, circle system, and ventilator diaphragm emitted traces of volatile anesthetics.
Conclusion:
In elective cases, presumably safe concentrations can be obtained by a 10-minute flush at ≥10 L/minute between removal and replacement all components of the airway circuit. For emergencies, we recommend using an activated charcoal filter.
More Related Videos
04:56In Vitro Method to Control Concentrations of Halogenated Gases in Cultured Alveolar Epithelial Cells
Published on: October 23, 2018
05:39Author Spotlight: A Non-Intubated Video-Assisted Thoracoscopic Surgery with Multimodal Analgesia and Sevoflurane Inhalation Anesthesia
Published on: May 26, 2023
Related Concept Videos
Prevention of Further Absorption of Poison
Inhalational Anesthetics: Overview
Parenteral Anesthetics: Overview
Depolarizing Blockers: Pharmocokinetics
Suctioning the Nasopharyngeal Airway
Equipment Required
Local Anesthetics: Common Agents and Their Applications
Cocaine is an ester of benzoic acid and methylecgogine. It is used to anesthetize and vasoconstrict locally. Currently, it is used primarily for topical applications. It is beneficial for surgeries on the upper respiratory tract, providing anesthesia and shrinking the mucosa. Cocaine in the form of cocaine hydrochloride is...