Effects of high relative humidity and dry purging on VOCs obtained during breath sampling on common sorbent tubes
Maxim Wilkinson1, Iain R White, Royston Goodacre
1Division of Infection, Immunity and Respiratory Medicine, School of Biological Sciences, Faculty of Biology, Medicine and Health, University of Manchester, Manchester, United Kingdom.
Abstract:
Offline breath analysis by thermal desorption gas chromatography mass spectrometry (TD-GC-MS) requires the use of sorbent traps to concentrate and store volatile compounds. The selection of which sorbent to use and best practices for managing high relative humidity are important considerations to allow for reproducible, untargeted, biomarker discovery in water saturated breath samples. This work aims to assess three commonly used sorbent materials for their use in breath volatile sampling and determine how the high relative humidity inherent in such samples effects the capture of volatile compounds of interest. TenaxGR, TenaxTA/Carbograph1TD and TenaxTA/Carbograph5TD tubes were selected as they are the most commonly used sorbents in the breath sampling literature. The recovery of 29 compounds in a standard mix loaded using high humidity gas was tested for each sorbent and compared to loading in dry gas. Water retention and dry purge rates were determined for each sorbent for 500 ml and 1000 ml breath collections. Finally, breath samples were collected simultaneously on to each sorbent type using the ReCIVA and analysed by TD-GC-MS. All three sorbents exhibited acceptable reproducibility when loaded with the standard mix in dry gas (RSD < 10%). Loading the standard mix in humid gas led to reduced recovery of compounds based on their chemical properties. Dry purging performance for each sorbent material was assessed and was shown to be 1.14, 1.13 and 0.89 mg H2O min-1 for TenaxGR, TenaxTA/Carbograph1TD and TenaxTA/Carbograph5TD respectively when flushed with 50 ml min-1 of N2. A comparison of breath profiles on different sorbents showed differences in background artefacts (sulfur dioxide, cyclopenten-1-one and 3-nonene) and endogenous breath compounds (2-methyl-furan and furfural). This work demonstrates that high relative humidity during sampling reduces the ability of sorbent tubes to capture volatile compounds and could impact method detection limits during breath sampling. Sufficient water to impair accurate analysis was retained on all tubes. Minimal differences were observed between sorbent materials when used to sample breath, however, suggestions are provided for sorbent selection for future studies.
Related Concept Videos
Volatilization
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...
Sputum Studies II: Culture and Sensitivity
Sputum culture and sensitivity is a medical procedure used to diagnose bacterial infections in the respiratory tract and select the most appropriate antibiotics for treatment. This process involves analyzing sputum samples of thick and opaque secretions produced in the lungs and airways. These samples are collected from patients and then sent to the laboratory for analysis.
The test can identify various pathogens responsible for respiratory infections, including Streptococcus,...