Myotonometry for the evaluation of Achilles tendon mechanical properties: a reliability and construct validity study
Alessandro Schneebeli1,2, Deborah Falla2, Ron Clijsen1,3
1Rehabilitation Research Laboratory 2rLab, Department of Business Economics, Health and Social Care, University of Applied Sciences and Arts of Southern Switzerland, Manno/Landquart, Switzerland.
BMJ Open Sport & Exercise Medicine
|March 11, 2020
Summary
The MyotonPRO device demonstrates high reliability for assessing Achilles tendon stiffness in various positions and contraction levels. This makes it a dependable tool for evaluating tendon mechanical properties.
Area of Science:
- Biomechanics
- Musculoskeletal research
- Medical device technology
Background:
- Assessing Achilles tendon stiffness is crucial for understanding lower limb biomechanics and injury mechanisms.
- Objective measurement tools are needed to complement traditional clinical assessments.
Purpose of the Study:
- To evaluate the intra-rater and inter-rater reliability of the MyotonPRO device.
- To assess the construct validity of the MyotonPRO for measuring Achilles tendon stiffness.
Main Methods:
- A reliability and construct validity study was conducted with 40 healthy participants.
- The MyotonPRO was used by two raters on two separate occasions, assessing the Achilles tendon in relaxed, 0° plantarflexion, and standing positions.
- Measurements were taken during isometric contractions (0-3 kg), with reliability analyzed using intraclass correlation coefficients (ICC).
Main Results:
- Very high intra-rater reliability (ICC2,k 0.87-0.98) and high to very high inter-rater reliability (ICC2,k 0.76-0.86) were observed.
- High to very high inter-session reliability (ICC2,k 0.70-0.89) was also found.
- Moderate reliability was noted for specific conditions, such as 0.5 kg contraction (ICC2,k 0.54-0.59) and the standing position (ICC2,k 0.56).
Conclusions:
- The MyotonPRO is a reliable and valid instrument for evaluating Achilles tendon stiffness across different contraction levels and positions.
- The device shows promise for clinical implementation in assessing tendon tissue mechanical properties.