Influence of Congruency Design on the Contact Stress of a Novel Hinged Knee Prosthesis Using Finite Element Analysis
Jing-Yu Zhang1,2, Dong-Mu Tian3, Zhi-Peng Ren1
1Department of Bone and Soft Tissue Oncology, Tianjin Hospital, Tianjin, China.
Objectives:
To investigate the contact stress and the contact area o tibial inserts and bushings with respect to different congruency designs in a spherical center axis and rotating bearing hinge knee prosthesis under gait cycle loading conditions using finite element analysis.
Methods:
Nine prostheses with different congruency (different degrees of tibiofemoral conformity and different distances between the spherical center and the bushing) designs were developed with the same femoral and tibial components. The models were transferred to finite element software. The peak contact stresses and contact areas on tibial inserts and bushings under the gait cycle loading conditions were investigated and compared.
Results:
For tibial insert, the peak contact stress was the highest in the low conformity-long group (61.4486 MPa), and it was 1.88 times higher than that in the group with the lowest stress (moderate conformity-short group, 32.754 MPa). The contact area was the largest in the low conformity-long group (420.485 mm2 ), and it was 1.19 times larger than that in the group with the smallest area (moderate conformity-middle group, 352.332 mm2 ). For bushing, the peak contact stress was the highest in the high conformity-long group (72.8093 MPa), and it was 3.21 times higher than that in the group with the lowest stress (high conformity-short group, 22.6928 MPa). The contact area was the largest in the low conformity-short group (2.41 mm2 ), and it was 2.27 times larger than that in the group with the smallest area (high conformity-middle group, 1.063 mm2 ).
Conclusion:
The results of our study showed that the congruency of the tibiofemoral surface and bushing surface should be considered carefully in the design of the spherical center axis and rotating bearing hinge knee prosthesis. Different levels of contact performance were observed with different congruency designs. In addition, the influence of contact stress and contact area on the polyethylene wear of rotating hinge knee prostheses should be confirmed with additional laboratory tests.
Related Concept Videos
Bending of Members Made of Several Materials
Hooke's Law determines stress in each material, stating that stress is proportional to strain but varies due to each material's...
Generalized Hooke's Law
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to its distance...
Stress Concentrations
Stress Concentrations
The stress...