The Precision of Propex Pixi with Different Instruments and Coronal Preflaring Procedures
Inês Ferreira1, Ana Cristina Braga2, Irene Pina-Vaz1
1Department of Endodontics, Faculty of Dental Medicine of University of Porto, Porto, Portugal.
European Endodontic Journal
|March 13, 2020
Summary
Coronal preflaring improves the accuracy of Propex Pixi electronic apex locators. This enhancement occurred regardless of the file alloy or apical fit, making Propex Pixi a precise instrument for root canal length determination.
Area of Science:
- Endodontics
- Dental materials science
Background:
- Accurate determination of root canal working length is crucial for successful endodontic treatment.
- Electronic apex locators (EALs) are widely used to measure canal length, but their accuracy can be influenced by various factors.
- The Propex Pixi is an electronic apex locator whose precision requires evaluation under different clinical conditions.
Purpose of the Study:
- To assess the accuracy of the Propex Pixi electronic apex locator.
- To investigate the influence of coronal preflaring on Propex Pixi accuracy.
- To evaluate the impact of different file alloys (stainless steel and nickel-titanium) and sizes on the instrument's performance.
Main Methods:
- Extracted human teeth were used to establish actual working length (WL) with a dental microscope.
- Propex Pixi was employed to measure electronic length (EL) to the root apex.
- Measurements were taken using various file alloys and sizes, both before and after coronal preflaring.
- Statistical analysis utilized factorial analysis of variance.
Main Results:
- Coronal preflaring significantly improved the accuracy of electronic length measurements (EL closer to WL).
- A strong intraclass correlation was found between electronic and actual working lengths.
- No significant differences in accuracy were observed based on file alloy or size.
Conclusions:
- The Propex Pixi demonstrated adequate precision for determining root canal length.
- Coronal preflaring procedures enhance the accuracy of the Propex Pixi, irrespective of the instrument type or apical fit.
- These findings support the routine use of coronal preflaring to optimize EAL performance.