CHARACTERIZATION OF TWO CT SYSTEMS USING A CHANNELIZED HOTELLING OBSERVER AND NPS METRIC
O Ortenzia1, A D'Alessio2, L Noferini3
1Department of Medical Physics, University Hospital of Parma, Italy.
Abstract:
We investigated the performances of two computed tomography (CT) systems produced by the same manufacturers (Somatom Flash and Edge Siemens) with different detector technologies (Ultrafast Ceramic and Stellar) and different generation of iterative reconstruction (IR) algorithms (SAFIRE and ADMIRE). A homemade phantom was scanned and the images were reconstructed with filtered back-projection (FBP) and IR algorithms. In terms of image quality, the performances of the systems were checked using the low-contrast detectability, evaluated by a Channelized Hotelling Observer (CHO), and the noise power spectrum (NPS). The analysis with CHO showed the best performance of Edge respect to Flash system for both FBP and IR algorithms. This better behavior, which reaches 20%, has been ascribed to the Stellar detector. From the NPS analysis, the noise reduction due to Stellar detector was 57%, moreover ADMIRE algorithm preserves a more traditional CT image texture appearance versus SAFIRE due to a lower NPS peak shift.
Related Concept Videos
Receiver Operating Characteristic Plot
Feedback control systems
Linear feedback systems are theoretical models that simplify analysis and design. These systems operate under the principle that their output is directly proportional to their input within certain ranges. For instance, an amplifier in a control system behaves linearly as long as the input signal remains within a specific range. However, most physical systems exhibit inherent nonlinearity...
Linear time-invariant Systems
The input-output behavior of an LTI system can be fully defined by its response to an impulsive excitation at its input. Once this impulse response is known, the system's reaction to any other input can be...
Classification of Systems-I
Homogeneity dictates that if an input x(t) is multiplied by a constant c, the output y(t) is multiplied by the same constant. Mathematically, this is expressed as:
Classification of Systems-II
Energy and Power Signals