Related Experiment Video
Updated: Dec 26, 2025

Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
A Carotid Doppler Patch Accurately Tracks Stroke Volume Changes During a Preload-Modifying Maneuver in Healthy
Jon-Émile S Kenny1, Igor Barjaktarevic2, Andrew M Eibl1
1Health Sciences North Research Institute, Sudbury ON, Canada.
Objectives:
Detecting instantaneous stroke volume change in response to altered cardiac preload is the physiologic foundation for determining preload responsiveness.
Design:
Proof-of-concept physiology study.
Setting:
Research simulation laboratory.
Subjects:
Twelve healthy volunteers.
Interventions:
A wireless continuous wave Doppler ultrasound patch was used to measure carotid velocity time integral and carotid corrected flow time during a squat maneuver. The Doppler patch measurements were compared with simultaneous stroke volume measurements obtained from a noninvasive cardiac output monitor.
Measurements And Main Results:
From stand to squat, stroke volume increased by 24% while carotid velocity time integral and carotid corrected flow time increased by 32% and 9%, respectively. From squat to stand, stroke volume decreased by 13%, while carotid velocity time integral and carotid corrected flow time decreased by 24% and 10%, respectively. Both changes in carotid velocity time integral and corrected flow time were closely correlated with changes in stroke volume (r 2 = 0.81 and 0.62, respectively). The four-quadrant plot found a 100% concordance rate between changes in stroke volume and both changes in carotid velocity time integral and changes in corrected flow time. A change in carotid velocity time integral greater than 15% predicted a change in stroke volume greater than 10% with a sensitivity of 95% and a specificity of 92%. A change in carotid corrected flow time greater than 4% predicted a change in stroke volume greater than 10% with a sensitivity of 90% and a specificity of 92%.
Conclusions:
In healthy volunteers, both carotid velocity time integral and carotid corrected flow time measured by a wireless Doppler patch were useful to track changes in stroke volume induced by a preload-modifying maneuver with high sensitivity and specificity.
More Related Videos
12:15Optimized System for Cerebral Perfusion Monitoring in the Rat Stroke Model of Intraluminal Middle Cerebral Artery Occlusion
Published on: February 17, 2013
08:42Ultrasound Assessment of Endothelial-Dependent Flow-Mediated Vasodilation of the Brachial Artery in Clinical Research
Published on: October 22, 2014
Related Concept Videos
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...