Contribution of usage to endoscope working channel damage and bacterial contamination
L C S Santos1, F Parvin2, A Huizer-Pajkos2
1Surgical Infection Research Group, Faculty of Medicine and Health Sciences, Macquarie University, Sydney, Australia; Ribeirao Preto Nursing School, Sao Paulo University, Ribeirao Preto, Sao Paulo, Brazil.
Background:
Biofilm formation has been shown to be associated with damaged areas of endoscope channels. It was hypothesized that the passage of instruments and brushes through endoscope channels during procedures and cleaning contributes to channel damage, bacterial attachment and biofilm formation.
Aim:
To compare surface roughness and bacterial attachment in used and new endoscope channels in vivo and in vitro.
Methods:
Surface roughness of 10 clinically used (retired) and seven new colonoscope biopsy channels was analysed by a surface profiler. For the in-vitro study, a flexible endoscope biopsy forceps was passed repeatedly through a curved 3.0-mm-diameter Teflon tube 100, 200 and 500 times. Atomic force microscopy was used to determine the degree of inner surface damage. The number of Escherichia coli or Enterococcus faecium attached to the inner surface of the new Teflon tube and the tube with 500 forceps passes in 1 h at 37oC was determined by culture.
Results:
The average surface roughness of the used biopsy channels was found to be 1.5 times greater than that of the new biopsy channels (P=0.03). Surface roughness of Teflon tubes with 100, 200 and 500 forceps passes was 1.05-, 1.12- and 3.2-fold (P=0.025) greater than the roughness of the new Teflon tubes, respectively. The number of E. coli and E. faecium attached to Teflon tubes with 500 forceps passes was 2.9-fold (P=0.021) and 4.3-fold (P=0.004) higher compared with the number of E. coli and E. faecium attached to the new Teflon tubes, respectively.
Conclusion:
An association was found between endoscope usage with damage to the biopsy channel and increased bacterial attachment.
More Related Videos
15:49Flexible Colonoscopy in Mice to Evaluate the Severity of Colitis and Colorectal Tumors Using a Validated Endoscopic Scoring System
Published on: October 16, 2013
09:50Induction of Intestinal Graft-versus-host Disease and Its Mini-endoscopic Assessment in Live Mice
Published on: February 11, 2019
Related Concept Videos
Endoscopic Procedures III: Video Capsule Endoscopy
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Endoscopic Procedures V: ERCP
Patient...