End-to-end test to evaluate the comprehensive geometric accuracy of a proton rotating gantry using a cone-shaped
Takahiro Kato1,2, Yuhei Yamazaki3, Ryohei Kato3
1Preparing Section for New Faculty of Medical Science, Fukushima Medical University, 1 Hikariga-oka, Fukushima, 960-1295, Japan. kato.newjapan@gmail.com.
Abstract:
In this study, we aim to evaluate the comprehensive geometric accuracy of proton rotating gantries by performing an end-to-end test using a cone-shaped scintillator screen detector, known as XRV-124. The XRV-124 comprises a cone-shaped sheet-like scintillator and charge-coupled device camera that detects the scintillation light. First, the results of the Winston-Lutz and end-to-end XRV-124 tests performed on a conventional linear accelerator were compared to confirm the reliability of the XRV-124, and the snout position dependency of the geometric accuracy was evaluated for the proton rotating gantry as a pre-verification process. Thereafter, an end-to-end test including computed tomography imaging and irradiation in 30° steps from 0° to 330° for two proton rotating gantries, which have the same specifications, was performed. The results of the pre-verification indicated that sufficient accuracy was obtained for the end-to-end test of the proton rotating gantry. The end-to-end test results showed a peak-to-peak deviation of up to 2 mm for some of the coordinate axes. The two gantries exhibited almost similar results in terms of the absolute quantity; however, a few trends were different. Thus, the beam axis deviations were confirmed to be within the safety margin, as expected in clinical practice. Based on the results of this study, the XRV-124 can be used as a comprehensive end-to-end constancy test tool, as it enables a comparative verification of multiple rotating gantries and geometric accuracy verification of different treatment modalities.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020