Development of an ergonomic four-finger-push manual pipette design
Eunsik Kim1, Faisal Aqlan2, Andris Freivalds3
1Mechanical, Automotive, and Materials Engineering, University of Windsor, 401 Sunset Ave, Windsor, ON, N9B 3P4, Canada.
A new four-finger pipette design significantly reduces muscle exertion and awkward wrist posture compared to traditional pipettes. Most users preferred the ergonomic design, suggesting it can lower risks of work-related musculoskeletal disorders (WMSDs).
Area of Science:
- Ergonomics
- Human Factors Engineering
- Occupational Health
Background:
- Traditional pipettes often require awkward hand and wrist postures, contributing to discomfort and potential work-related musculoskeletal disorders (WMSDs).
- Existing pipette designs may not adequately consider hand anatomy and anthropometry, leading to suboptimal user experience and increased risk factors.
Purpose of the Study:
- To develop and evaluate novel pipette designs operated by four fingers, incorporating hand anatomy and anthropometry.
- To compare the ergonomic performance of the new pipette designs against traditional thumb-push pipettes.
Main Methods:
- Evaluated muscle activity, wrist posture (flexion, radial/ulnar deviation), and subjective discomfort ratings for upper extremities.
- Conducted user preference tests comparing the four-finger pipette designs with a traditional thumb-push pipette.
Main Results:
- The four-finger ergonomic pipette design demonstrated significant reductions in muscle exertion (25% aspiration, 35% dispensing) and awkward wrist posture (33% reduction).
- Participants reported decreased perceived discomfort in the wrist, hand, and lower arm when using the new designs.
- A strong user preference (9 out of 10 participants) was observed for the new pipette designs over the traditional model.
Conclusions:
- The developed four-finger pipette design effectively mitigates ergonomic risk factors associated with pipette use.
- This ergonomic innovation has the potential to reduce pain and the incidence of WMSDs in laboratory and clinical settings.
- The findings support the adoption of anthropometrically-informed designs for laboratory tools to enhance user well-being.