Related Experiment Video
Updated: Dec 26, 2025

Measurements of Soil Carbon by Neutron-Gamma Analysis in Static and Scanning Modes
Published on: August 24, 2017
Calibration of a CeBr3 based γ-spectrometer for onsite and laboratory radioactivity measurements in sediment samples
E G Androulakaki1, C Tsabaris1, F Maragkos2
1Hellenic Centre for Marine Research, Institute of Oceanography, P.O. Box 712, GR-19013, Anavyssos, Greece.
Abstract:
The calibration of an underwater spectrometer, consisting of a CeBr3 crystal, is performed for sediment sample measurements, via experimental data and Monte Carlo simulations. The system was implemented for activity concentration measurements in marine sediment samples in the laboratory. The results were compared with corresponding high resolution measurements and showed that the system provides accurate results, but relatively high uncertainties. Different measurement geometries are theoretically studied via Monte Carlo simulations in order to improve the system performance for such measurements.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...