Comparison of a Mid Scala and a Perimodiolar Electrode in Adults: Performance, Impedances, and Psychophysics
Rolf-Dieter Battmer1, Sandra Scholz2, Dzemal Gazibegovic3
1Department Otolaryngology at ukb, Berlin.
Objectives:
The HiFocus Mid-Scala electrode array (HFms) is designed to sit within the scala tympani without touching either the lateral wall or the modiolus. The aim of this study was to compare the HFms to the Helix perimodiolar electrode array.
Method:
Two groups of recipients with Helix (n = 22 ears) and HFms (n = 29 ears) electrode arrays were retrospectively identified and matched by age at implantation and duration of severe to profound deafness. Most comfortable listening levels (M), impedances, Freiburger Monosyllables in quiet, and Oldenburg sentences in adaptive noise were compared at 3, 6, and 12 months postimplant.
Results:
Median scores for monosyllables in quiet for the HFms group were significantly better than the Helix group at each test interval (p < 0.05). Speech perception in quiet also significantly improved from 3 to 12 months for both groups (p < 0.001). There was no significant difference between the groups for speech in noise. Impedances were significantly lower for the HFms group at 12 months (p < 0.05) except at the basal end and M levels were generally higher.
Conclusions:
The HFms group had better median performance for monosyllables in quiet than the Helix group at each test interval, although performance in noise was similar. For speech in noise, the HFms group appear to reach optimum performance quicker than the Helix group. Impedances were lower in the HFms group across the array, other than at the most basal end, and support our hypothesis that the HFms assumes a more lateral position within the cochlea than the Helix electrode, although our article did not include imaging data.
More Related Videos
08:24The Joint Effect of Social Comparison and Social Distance on Evaluation of Intertemporal Choice Outcomes in Event-related Potential Studies
Published on: August 25, 2023
09:27A Method for Systematic Electrochemical and Electrophysiological Evaluation of Neural Recording Electrodes
Published on: March 3, 2014
Related Concept Videos
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Bode Plots Construction