Purpose:
In context of head-and-neck radiotherapy, this study aims to compare MR image quality according to diagnostic (DIAG) and radiotherapy (RT) setups; and to optimise an MRI-protocol (including 3D T1 and T2-weighted sequences) for dose-planning (based on pseudo-CT generation).
Materials And Methods:
To compare DIAG and RT setups, signal-to-noise-ratio (SNR) and percentage-image-uniformity (PIU) were computed on T1 images of phantoms and volunteers. Influence of the sample conductivity on SNR was quantified using homemade phantoms. To obtain reliable T1 and T2 images for RT-planning, an experimental design was performed on volunteers by using SNR, contrast-to-noise-ratio (CNR) and mean-opinion-score (MOS). Further, pseudo-CTs were generated from 8 patients T2 images with a state-of-art deep-learning method. These pseudo-CTs were evaluated by mean-absolute-error (MAE) and mean-error (ME).
Results:
SNR was higher for DIAG-setup compared to RT-setup (SNR-ratio=1.3). A clear influence of the conductivity on SNR was observed. PIU was higher for DIAG-setup (38.8%) compared to RT-setup (33.5%). Regarding the protocol optimisation, SNR, CNR, and MOS were 20.6, 6.16, and 3.91 for the optimal T1 sequence. For the optimal T2 sequence, SNR, CNR and MOS were 25.6, 44.46 and 4.0. In the whole head-and-neck area, the mean MAE and ME of the pseudo-CTs were 82.8 and -3.9 HU.
Conclusion:
We quantified the image quality decrease induces by using an RT-setup for head-and-neck radiotherapy. To compensate this decrease, an MRI protocol was optimised by using an experimental design. This protocol of 15minutes provides accurate images which could be used for MRI-dose-planning in clinical practice.