Variability in expiratory flow requirements among oscillatory positive expiratory pressure
Sherwin E Morgan1, Steven Mosakowski1, Brenda L Giles2
1Department of Respiratory Therapy, 5841 S. Maryland Ave MC4028, University of Chicago, Chicago, IL 60637.
Introduction:
Oscillatory positive expiratory pressure (OPEP) devices facilitate secretion clearance by generating positive end expiratory pressure. However, different device designs may produce different levels of expiratory pressure with the same expiratory flow rate. We bench tested four devices to determine the relationship between expiratory flow and expiratory pressure in each.
Methods:
A bench model was created to test the gas flow rates required by different OPEP devices to generate target expiratory pressure. Four different devices were tested: Acapella® (DH Green, Smiths Medical), AerobiKa® (Monaghan Medical Corporation), VibraPEP® (Curaplex), and vPEP™ (D R Burton Healthcare). Each OPEP device was tested to determine the expiratory flow needed to generate expiratory pressure thresholds considered appropriate for OPEP therapy.
Results:
The expiratory flow required to generate the same expiratory pressure thresholds varied considerably among devices. Valved OPEP devices such as the VibraPEP required less flow than mechanical devices such as the vPEP, Aerobika, and Acapella.
Discussion:
In this bench test of OPEP devices, we found considerable variability in expiratory flow requirements needed to generate an expiratory pressure of >10 cm H2O. Our finding suggests that smaller patients or those with limited expiratory airflow due to diseases such as COPD, obesity, chronic congestive heart failure, and restrictive lung disease may have better results when matched to OPEP devices requiring less expiratory airflow.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...