Related Experiment Video For Vibratron II
Updated: Dec 26, 2025

A Simple Non-invasive Method for Temporary Knockdown of Upper Limb Proprioception
Published on: March 3, 2018
Test-Retest Reliability of Vibration Perception Threshold Test in People with Type 2 Diabetes Mellitus
Francisco Javier Domínguez-Muñoz1, José Carmelo Adsuar2, Santos Villafaina1
1Physical Activity and Quality of Life Research Group (AFYCAV), Faculty of Sport Science, University of Extremadura, 10003 Cáceres, Spain.
Background:
Diabetes mellitus is a chronic disease characterized by fasting hyperglycemia. It affects approximately 415 million people worldwide and involves a variety of complications. One of them is the loss of sensitivity to peripheral vibration.
Objective:
Our study aims to discover the test-retest reliability of a procedure for assessing vibration sensitivity in people with type 2 diabetes mellitus.
Methodology:
90 people with type 2 diabetes mellitus (56 men and 34 women) performed the vibration perception threshold (VPT) test using the Vibratron II device. A re-test was completed seven days after the first reading.
Results:
The relative reliability of the VPT test result is excellent (intraclass correlation coefficient = 0.96). The same applies to gender and obesity subgroups. Regarding absolute reliability, the standard error of measurement is 8.99%, and the small real difference is 24.94%.
Conclusions:
The relative and absolute reliability results of the vibration perception threshold in people with type 2 diabetes mellitus offer excellent results.