Site qualification and clinical interpretation standards for 99mTc-SPECT perfusion imaging in a multi-center study of
Jennifer M Renaud1, Manuja Premaratne1,2,3,4, Marie-Claude Villeneuve5
1Division of Cardiology and National Cardiac PET Centre, University of Ottawa Heart Institute, 40 Ruskin Street, Ottawa, ON, K1Y 4W7, Canada.
Background:
Qualification and interpretation standards are essential for establishing 99mTc-SPECT MPI accuracy vs. alternative modalities.
Methods:
Rest-stress 99mTc-SPECT phantom scans were acquired on 35 cameras. LV defects were quantified with summed stress (SSS) and difference scores (SDS) at 2 core labs. SDS ≥ 2 in the right coronary artery (RCA) was the qualifying standard. Twenty rest (R)-stress (S) patient images were acquired on qualified cameras and interpreted by core labs. Global scoring differences > 3 between labs or discordant clinical interpretations underwent review. Scoring, interpretation, image quality, and diagnostic parameter agreement were assessed.
Results:
Phantom scans: visual scoring confirmed RCA-ischemia on all cameras. Regional SSS, SDS agreement was moderate to very good: ICC-r = 0.57, 0.84. Patient scans: 90% of global SSS, 85% of SDS differences were ≤ 3. Regional SSS, SDS agreement: ICC-r = 0.87, 0.86, and global abnormal (SSS ≥ 4) and ischemic (SDS ≥ 2) interpretation: ICC-r = 0.90 were excellent. Clinical interpretation agreement was 100% following review. Image quality agreement was 70%. Automated metrics also agreed: ischemic total perfusion deficit ICC-r = 0.75, reversible perfusion defect, transient ischemic dilation, and S-R LV ejection fraction ICC-r ≥ 0.90.
Conclusion:
Quantitative scoring and interpretation of scans were highly repeatable with site qualification and clinical interpretation standardization, indicating that dual-core lab interpretation is appropriate to determine 99mTc-SPECT MPI accuracy.