Development of a simple high performance liquid chromatography (HPLC)/evaporative light scattering detector (ELSD)
Bikash Mondal1, Mahesh Kote1, Chandrakant Lunagariya1
1Research and Development, Akorn Pharmaceuticals, 50 Lake View Parkway, Suite#112, Vernon Hills, IL, 60061, United States.
Abstract:
The amount of polysorbate 80 in pharmaceutical formulations affects the product quality and efficacy. A reliable test method is required to quantify the amount of Polysorbate 80 present in the drug product formulations. The test method for the determination of Polysorbate 80 may be used during process development and final product quality assessment. A simple, fast and efficient quantitative method, making use of HPLC-ELSD and a C18 column without sample pretreatment was developed. The developed method demonstrated specificity to polysorbate 80 with high precision as indicated by percent relative standard deviation (%RSD) of 3.0% for six determinations. The accuracy of this method for the determination of polysorbate 80 in a pharmaceutical formulation was demonstrated with an overall recovery of 94.9%.
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Instrumentation
High-Performance Liquid Chromatography: Elution Process
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...