Mobile-bearing insert used with total knee arthroplasty does not rotate on the tibial tray during a squatting
Kenji Hoshi1, Goro Watanabe1, Yasuo Kurose2
1Graduate School of Medical Technology and Health Welfare Sciences, Hiroshima International University, 555-36 Kurose-Gakuendai, Bldg3, Rm3807, Higashi-Hiroshima, Hiroshima, 739-2695, Japan.
Background:
Total knee arthroplasty (TKA) is commonly performed around the world. Implant designs include fixed-bearing and mobile-bearing. Mobile-bearing design was developed as a rotating platform that allows axial rotation of the insert around the longitudinal axis. This phenomenon may limit full exploitation of the characteristics of the mobile-bearing insert, which may cause wearing and reduce longevity. However, there is limited knowledge on rotational behavior of the polyethylene mobile-bearing insert under weight-bearing conditions. We aimed at determining the rotational motion of each component at full extension and flexed positions during a squatting activity after TKA.
Methods:
This study was a cross-sectional study (level 4) involving patients with severe knee osteoarthritis scheduled to receive TKA. We examined 13 knees of 11 patients after mobile-bearing TKA (NexGen LPS-Flex, Zimmer Inc.) at 10 weeks and 1 year postoperatively. Four identical metallic beads were embedded into the insert. Wide-base squatting was chosen for analyses. Three-dimensional in vivo poses of the prostheses were created using a 3D-to-2D registration technique. During flexion, rotation of the femoral component relative to the insert (FEM/INS) and tibial component (FEM/TIB) as well as insert rotation relative to the tibial component (INS/TIB) were computed. Repeated measure 2-way ANOVA and post hoc test was used.
Results:
In the fully extended position, FEM/INS was significantly smaller than INS/TIB both at 10 weeks (- 0.3° vs. 6.3°, p = .013) and 1 year (- 0.8° vs. 4.9°, p = .011), respectively. During the squatting activity, rotation motions of FEM/TIB, FEM/INS, INS/TIB were 5.7°, 5.9°, and 1.8° at 10 weeks and 6.3°, 5.5°, and 1.6° at 1 year, respectively. Rotation motion of FEM/INS was significantly greater than that of INS/TIB at both 10 weeks (p < .001) and 1 year (p < .001).
Conclusions:
The mobile-bearing insert enhances the compatibility of FEM/INS in extension; the amount of INS/TIB rotation is significantly smaller than that of FEM/INS during a squatting activity. This information will inform surgeons to take caution to perform TKA with a fixed insert in which 6.3° of rotational offset would be added to the rotational alignment at FEM/INS at full extension.
Trial Registration:
UMIN-CTR, UMIN000024196. Retrospectively registered on 9 September 2016.
Related Concept Videos
Knee Joint
A total of seven ligaments support the knee joint. The patellar ligament, which is also attached to the quadriceps femoris...
Muscles that Move the Leg
Anterior Compartment
The quadriceps femoris, the most visible muscle of the anterior compartment, is integral for leg extension and thigh flexion. It is formed by merging four distinct muscles — the vastus lateralis, vastus medialis, vastus intermedius, and rectus femoris. The quadriceps tendon, a shared tendon of the four quadriceps muscles, is affixed...