Rapid inductively coupled plasma mass spectrometry method to determine iodine in amniotic fluid, breast milk and
Yutong Zou1, Danchen Wang1, Songlin Yu1
1Department of Clinical Laboratory, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Science, Beijing 100730, China.
Background:
Given that both insufficient and excess maternal iodine have adverse consequences such as poor cognitive performance, delayed physical development and increased fetal and infant mortality, the determination of maternal iodine status is very important. In this study, we established and verified a method involving inductively coupled plasma-mass spectrometry (ICP-MS) technology for the rapid determination of the amniotic fluid iodine concentration (AFIC), breast milk iodine concentration (BMIC) and cerebrospinal fluid (CSF) iodine concentration (CSFIC).
Method:
Amniotic fluid, breast milk and CSF were collected from residual samples at Peking Union Medical College Hospital (PUMCH). The linearity, detection limit, precision, recovery, carryover and matrix effect of the testing method using ICP-MS technology were thoroughly evaluated according to the EP-10-A2 evaluation protocol approved by the Clinical and Laboratory Standards Institute (CLSI) guidelines. Furthermore, we evaluated the AFIC, BMIC and CSFIC distributions among clinical patients from PUMCH.
Results:
The correlation coefficient (r) was higher than 0.99 (0.995-1.000). The limit of detection (LOD) was 0.233 µg/L, and the limit of quantification (LOQ) was 0.778 µg/L. For amniotic fluid, breast milk and CSF, the assay repeatabilities were 1.5%-1.8%, 1.9%-4.0% and 1.8%-4.0%, respectively, and the within-laboratory coefficient of variations (CV%) over five days were 3.3%-9.2%, 7.2%-8.0% and 3.2%-7.8%, respectively. The recovery rates ranged from 97.7% to 109.8%. Moreover, the median concentrations of iodine in the amniotic fluid, breast milk and CSF of the patients from PUMCH were 176.3 µg/L, 136.0 µg/L, and 81.8 µg/L, respectively.
Conclusion:
A rapid, stable and accurate method that incorporates ICP-MS technology for the determination of iodine concentration was established for amniotic fluid, breast milk and CSF in this study.
More Related Videos
10:14Synthesis and Purification of Iodoaziridines Involving Quantitative Selection of the Optimal Stationary Phase for Chromatography
Published on: May 16, 2014
07:28An Efficient Method for Selective Desalination of Radioactive Iodine Anions by Using Gold Nanoparticles-Embedded Membrane Filter
Published on: July 13, 2018
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Inductively Coupled Plasma Atomic Emission Spectroscopy: Principle
The ions and electrons produced interact with the fluctuating magnetic field created by a water-cooled...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Redox Titration: Iodimetry and Iodometry
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....