Patients With Modular-Neck Total Hip Arthroplasty: A Brief Five-Year Follow-Up Study

Daisuke Inoue1, Camilo Restrepo2, Blake Nourie2

  • 1Department of Orthopaedic Surgery, Rothman Institute, Thomas Jefferson University Hospital, Philadelphia, PA; Department of Orthopaedic Surgery, Graduate School of Medical Science, Kanazawa University, Kanazawa, Ishikawa, Japan.

Abstract

Insights

Midterm follow-up of modular-neck hip stems revealed a nearly doubled revision rate for corrosion compared to early results. Clinical symptoms, not ion levels, appear to be the primary indicator of failure in these hip implants.

Area of Science:

  • Orthopedic surgery
  • Biomaterials science
  • Medical device engineering

Background:

  • Previous early results showed a 13% revision rate for modular-neck stem corrosion within 2 years.
  • This study provides midterm follow-up data for these implants.

Purpose of the Study:

  • To update clinical findings on modular-neck hip stems with a midterm mean follow-up of 5 years.
  • To assess the incidence and characteristics of neck-stem corrosion and its impact on revision rates.

Main Methods:

  • Retrospective review of 186 modular-neck hips in 175 patients.
  • Mean follow-up of 60.1 months.
  • Evaluation included clinical findings, radiographs, advanced imaging (MRI, ultrasound), and serum ion levels (cobalt, chromium).

Main Results:

  • 41 hips (22.0%) required revision due to neck-stem corrosion.
  • Symptomatic patients consistently presented with groin pain and/or swelling.
  • Serum cobalt levels increased over time; imaging revealed fluid collections and soft tissue masses in symptomatic cases.
  • Survival rate was 72% at 7 years, with 14% of revised patients lacking positive imaging findings.

Conclusions:

  • The midterm revision rate for modular-neck stem corrosion nearly doubled compared to early findings.
  • Clinical symptoms are more indicative of failure than serum ion level monitoring.
  • This highlights potential long-term concerns with modular-neck hip implant designs regarding corrosion.