Prospective analysis of cultures from the Furlow insertion tool: a possible etiology for penile prosthesis infections
Faysal A Yafi1, James Furr2, Farouk M El-Khatib3
1University of California Irvine, Irvine, CA, USA. Fyafi@uci.edu.
Abstract:
The most dreaded complication of penile prosthesis (PP) implantation is device infection. We sought to assess whether inadequate cleaning and sterilization of the reusable Furlow inserter may represent one of the last etiologies of infection in PP patients. We performed a prospective analysis of cultures of the Furlow inserter used for PP surgeries from seven centers between May 1st and June 30th, 2019. Once the Furlow was received for surgery, the surgical team inspected the device for assembly status (disassembled or not) and the presence of visible stains, pieces of tissue or discoloration on either the interior of the barrel or the plunger. Swab aerobic and anaerobic bacterial and fungal cultures were then obtained from the internal component, after removal from the external component if assembled, and after introduction and immediate removal from the external component if disassembled. A total of 83 Furlow devices were cultured. Median age of surgical instrument was 4 years (2-10 years). Methods of sterilization included autoclave, wet autoclave, steam, and Sterad. Median time from sterilization was 3 days (1-22). On initial presentation, 79 devices were disassembled (95.1%) and 4 devices were still assembled (4.9%). Three external components were discolored (3.6%), while internal components demonstrated two stains (2.4%) indicative of improper cleaning which were thought to be residual blood products. Overall, 2/83 (2.4%) devices revealed positive swab cultures for Staphylococcus epidermidis. Swab cultures were negative for fungi and anaerobic bacteria. This patient cohort will continue to be followed to see if device infection occurs but it is unlikely to be meaningful since contaminated Furlows were discarded. Improper cleaning and/or sterilization of the Furlow Insertion Instrument may represent a source of infection for patients undergoing PP implantation.
Insights
Inadequate cleaning of the reusable Furlow inserter, used for penile prosthesis (PP) implantation, may cause device infections. Cultures revealed contamination in some devices, highlighting potential risks for patients.
Area of Science:
- Urology
- Surgical Infections
- Medical Device Sterilization
Background:
- Penile prosthesis (PP) implantation is associated with device infection, a significant complication.
- The reusable Furlow inserter is critical for PP placement.
- Potential sources of infection, including surgical instrument reprocessing, require investigation.
Purpose of the Study:
- To evaluate the reusable Furlow inserter as a potential source of infection in penile prosthesis (PP) patients.
- To assess the efficacy of cleaning and sterilization protocols for the Furlow inserter.
Main Methods:
- Prospective analysis of cultures from 83 Furlow inserters across seven centers.
- Inspection for visible contamination (stains, tissue) before culturing internal components.
- Aerobic, anaerobic, and fungal cultures obtained from disassembled and assembled inserters.
Main Results:
- Three external components (3.6%) and two internal components (2.4%) showed visible signs of improper cleaning.
- Two devices (2.4%) yielded positive cultures for Staphylococcus epidermidis.
- All cultures were negative for fungi and anaerobic bacteria.
Conclusions:
- Improper cleaning and/or sterilization of the Furlow Insertion Instrument may be a source of infection in penile prosthesis (PP) patients.
- Visible contamination on the inserter suggests potential breaches in reprocessing protocols.
- Further monitoring is needed to correlate instrument contamination with patient outcomes, though contaminated instruments were discarded.