Related Experiment Video
Updated: Dec 25, 2025

Vessel-sparing Excision and Primary Anastomosis
Published on: January 7, 2019
Factors Associated with Negative Patient Experiences with Essure Sterilization
Emilie V J van Limburg Stirum1, Nisse V Clark1, Alexis Lindsey1
1Division of Minimally Invasive Gynecologic Surgery, Brigham and Women's Hospital, Boston Massachusetts.
Study Objective:
The objective of the study was to identify factors associated with negative patient experiences with Essure.
Design:
This was a retrospective cohort study and follow-up survey.
Setting:
The study was conducted in an academic setting.
Patients:
Patients included women who had an Essure placed between 2002 and 2017.
Methods:
The hospitals' database was queried to identify subjects and charts were reviewed to confirm medical information. Subjects were invited by mail, e-mail, or phone call to participate a survey regarding symptoms and satisfaction with Essure. A comparison was made between women who reported a negative experience with Essure versus those who did not. A multivariable logistic regression analysis was performed to identify subject or procedural characteristics associated with any negative experience with Essure sterilization.
Results:
Two hundred eighty-four women underwent Essure sterilization between 2002 and 2017, 42.3% of whom responded to the follow-up survey. Satisfaction with Essure was reported by 61.9% of respondents. Thirty-three percent of the respondents have undergone removal or desire removal of the device. The most frequent symptoms attributed to Essure were pelvic pain, dyspareunia, and vaginal bleeding. Forty-eight percent of the respondents were identified as having any negative experience with Essure. Factors associated with negative experiences with Essure included young age at placement (odds ratio [OR] 0.86; 95% confidence interval [CI] 0.79-0.94; P < .001), high gravidity (OR 1.39; 95% CI 1.14-1.71; P = .002), and absent history of abdominal surgery (OR 0.35; 95% CI 0.12-1.00; P = .049).
Conclusion:
Young age at placement, high gravidity, and absent history of abdominal surgery are factors associated with negative patient experiences following Essure sterilization. A negative experience with Essure is common, although dissatisfaction with the device is not always attributable to symptoms. This information could be considered when counseling women who plan removal of Essure.
Implications Statement:
Our study provides new follow-up data with respect to hysteroscopic sterilization. This research is the first to examine any cause for negative patient experiences with Essure. Understanding factors associated with negative patient experiences could improve patient counseling regarding the extent to which symptoms could be attributed to Essure as well as counseling women who want to undergo removal of the device. These factors could in turn prove to be predictors for successful resolution of symptoms after removal of Essure. Results of this study could also be used for developing future hysteroscopic sterilization techniques.
More Related Videos
Related Concept Videos
Barriers to Effective Communication II
Cultural barriers:
Differences in values, beliefs, religion, knowledge, and tradition can significantly impact communication. Awareness of nonverbal cues is critical, especially when conversing with a patient from a different culture. What appears appropriate in one culture may be inappropriate in another.
Semantic barriers:
As a result of their tendency to use...
Factors Affecting the Risk of Infection
The integrity and count of the white blood cells help the body resist pathogens and fight infection. When impaired, it reduces the body's resistance to pathogens. The acidic pH levels of the gastrointestinal, genitourinary tracts, and skin...
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Ethical Issues
Ethical Concerns in Healthcare:
Spermatogenesis
Errors occurring during blood pressure monitoring
Several factors...