Reproducibility of a magnet-based jaw motion analysis system
International Journal of Computerized Dentistry
|March 25, 2020
Summary
The Dental Motion Decoder system (DMS-System) shows high reproducibility in vitro. In vivo measurements exhibit greater variability, suggesting reliable clinical performance for this magnetic field-based dental device.
Area of Science:
- Dental Technology
- Biomechanics
- Medical Device Evaluation
Background:
- The Dental Motion Decoder system (DMS-System) utilizes magnetic field technology to record mandible movements.
- Recorded data can be used for articulator programming or direct computer-aided design (CAD) processing.
Purpose of the Study:
- To assess the in vitro and in vivo reproducibility of the Dental Motion Decoder system (DMS-System).
Main Methods:
- Simulated protrusive and laterotrusive jaw movements in vitro (Group M) and in vivo (Group P).
- Measurements were taken with single or multiple reference point initializations.
- Evaluated reproducibility using standard deviations of sagittal condylar path inclination angle (sCPIA) and Bennett angle (BA).
Main Results:
- In vitro measurements (M1, M2) demonstrated significantly higher reproducibility (lower SDs) for sCPIA and BA compared to in vivo measurements (P1, P2).
- In vitro sCPIA SDs ranged from 0.08 to 0.26 degrees; BA SDs ranged from 0.06 to 0.11 degrees.
- In vivo sCPIA SDs ranged from 0.61 to 1.4 degrees; BA SDs ranged from 0.45 to 0.65 degrees.
Conclusions:
- The in vitro deviation of the DMS-System is smaller than the in vivo biologic variance.
- Reliable measurements using the DMS-System under clinical conditions can be assumed.