[Hamilton-Thorne Integrated Visual Optical System Ⅰ (IVOSⅠ) versus IVOS Ⅱ for human sperm analysis]
Hua Zhou1, Qing Li1, Li-Yuan Guo1
1Research Institute of Obstetrics and Gynecology / Key Lab of Reproduction and Genetics of Guangdong Higher Education Institutions, The Third Hospital of Guangzhou Medical University, Guangzhou, Guangdong 510150, China.
Objective:
To compare the computer-assisted sperm analysis (CASA) systems Hamilton-Thorne Integrated Visual Optical System Ⅰ (IVOSⅠ) and IVOS Ⅱ after verifying the performance of the latter so as to ensure the accuracy of the results of analysis.
Methods:
Based on the criteria established in the 5th edition of the WHO Laboratory Manual for the Examination and Processing of Human Semen (WHO 5th Ed), we compared the main semen parameters obtained from IVOSⅠ with those generated by IVOS Ⅱ, and examined the consistency between the results of the two sperm analyzers.
Results:
The linear relationship of the outlier test, bias estimation and scatter plot and the results of the outlier test of the two systems all met the requirements of comparison analysis and showed an obvious correlativity. The application scope of the results obtained from the apparatus indicated a reasonable value range, with r = 0.988 for sperm concentration, r = 0.975 for sperm progressive motility (PR), and r = 0.981 for total sperm motility. Evaluation of the acceptability of the predicted bias showed that the allowable total error (TEa) to be 6.67% with sperm concentration at 12 × 106 /ml and 2.34% with PR < 31%, their upper limit of the allowable error < 1/2. The results of IVOS Ⅱ conformed to the requirements of the WHO 5th Ed.
Conclusions:
The main parameters derived from IVOSⅠ and IVOS Ⅱ are comparable and consistent, indicating that both can be used for the examination of semen samples.
More Related Videos
08:06Author Spotlight: Evaluating the Impact of Immediate Partial Removal of Cumulus-Oocyte Complexes on Fertilization Efficiency and Embryo Quality
Published on: October 18, 2024
12:22Multimodal Volumetric Retinal Imaging by Oblique Scanning Laser Ophthalmoscopy oSLO and Optical Coherence Tomography OCT
Published on: August 4, 2018
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies II: Ultrasonography
Meiosis II
Total Internal Reflection Fluorescence Microscopy