Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Multicentre study: reliability and repeatability of Scheimpflug system measurement in keratoconus
Beatriz de Luis Eguileor1, Pedro Arriola-Villalobos2, Jose Ignacio Pijoan Zubizarreta3
1Ophthalmology, Hospital Universitario Cruces, Barakaldo, Spain beatrizdeluis@gmail.com.
This study found the Pentacam HR device offers reliable measurements for keratoconus diagnosis. However, repeatability limits indicate a need for new progression monitoring criteria based on actual system measurements.
Area of Science:
- Ophthalmology
- Corneal Imaging
- Refractive Surgery
Background:
- Keratoconus diagnosis and monitoring are crucial for preserving vision.
- Accurate and repeatable measurements are essential for effective patient management.
- The Pentacam HR is a widely used device for corneal tomography.
Purpose of the Study:
- To evaluate the repeatability and reliability of key tomographic parameters for keratoconus characterization using the Pentacam HR.
- To assess the impact of keratoconus severity on measurement repeatability.
Main Methods:
- Analysis of 230 eyes from 158 keratoconus patients.
- Five consecutive corneal tomography examinations per eye using Pentacam HR.
- Classification of eyes into three grades based on maximum posterior elevation (max_BFS_post).
- Calculation of intraclass correlation coefficients (ICCs) and repeatability limits for tomographic and aberrometric parameters.
Main Results:
- Excellent ICCs (0.9) were observed for all parameters, indicating high reliability.
- Repeatability limits varied: >0.5D for power, 20° for astigmatism axis, 10µm for thinnest cornea, >0.1µm for aberrometry, and >15° for coma axis.
- Repeatability generally increased with keratoconus severity, except for the coma axis.
Conclusions:
- The Pentacam HR demonstrates high reliability, supporting its use in keratoconus diagnosis.
- Repeatability limits highlight the need for updated criteria for monitoring keratoconus progression.
- Future monitoring strategies should incorporate the practical measurement capabilities of the Pentacam HR.