Related Experiment Video
Updated: Dec 25, 2025

Scalable Fabrication of Stretchable, Dual Channel, Microfluidic Organ Chips
Published on: October 20, 2018
Fabrication of immobilized enzyme reactors with pillar arrays into polydimethylsiloxane microchip
Cynthia Nagy1, Adam Kecskemeti1, Attila Gaspar1
1Department of Inorganic and Analytical Chemistry, University of Debrecen, Egyetem ter 1., Debrecen, 4032, Hungary.
Abstract:
This paper demonstrates the design, efficiency and applicability of a simple and inexpensive microfluidic immobilized enzymatic reactor (IMER) for rapid protein digestion. The high surface-to-volume ratio (S/V) of the reactor was achieved by forming pillars in the channel. It was found that pillar arrays including dimensions of 40 μm × 40 μm as pillar diameter and interpillar distance can provide both relatively high S/V and flow rate in the PDMS chip, the fabrication of which was performed by means of soft lithography using average research laboratory infrastructure. CZE peptide maps of IMER-based digestions were compared to peptide maps obtained from standard in-solution digestion of proteins. The peak patterns of the electropherograms and the identified proteins were similar, however, digestion with the IMER requires less than 10 min, while in-solution digestion takes 16 h.