Related Experiment Video
Updated: Dec 25, 2025

PET and MRI Guided Irradiation of a Glioblastoma Rat Model Using a Micro-irradiator
Published on: December 28, 2017
High-precision image-guided proton irradiation of mouse brain sub-volumes
Theresa Suckert1, Johannes Müller2, Elke Beyreuther3
1OncoRay - National Center for Radiation Research in Oncology, Faculty of Medicine and University Hospital Carl Gustav Carus, Technische Universität Dresden, Germany; German Cancer Consortium (DKTK), Partner Site Dresden, and German Cancer Research Center (DKFZ), Germany.
Background And Purpose:
Proton radiotherapy offers the potential to reduce normal tissue toxicity. However, clinical safety margins, range uncertainties, and varying relative biological effectiveness (RBE) may result in a critical dose in tumor-surrounding normal tissue. To assess potential adverse effects in preclinical studies, image-guided proton mouse brain irradiation and analysis of DNA damage repair was established.
Material And Methods:
We designed and characterized a setup to shape proton beams with 7 mm range in water and 3 mm in diameter and commissioned a Monte Carlo model for in vivo dose simulation. Cone-beam computed tomography and orthogonal X-ray imaging were used to delineate the right hippocampus and position the mice. The brains of three C3H/HeNRj mice were irradiated with 8 Gy and excised 30 min later. Initial DNA double-strand breaks were visualized by staining brain sections for cell nuclei and γH2AX. Imaged sections were analyzed with an automated and validated processing pipeline to provide a quantitative, spatially resolved radiation damage indicator.
Results:
The analyzed DNA damage pattern clearly visualized the radiation effect in the mouse brains and could be mapped to the simulated dose distribution. The proton beam passed the right hippocampus and stopped in the central brain region for all evaluated mice.
Conclusion:
We established image-guided proton irradiation of mouse brains. The clinically oriented workflow facilitates (back-) translational studies. Geometric accuracy, detailed Monte Carlo dose simulations, and cell-based assessment enable a biologically and spatially resolved analysis of radiation response and RBE.
More Related Videos
08:02A High-Throughput Image-Guided Stereotactic Neuronavigation and Focused Ultrasound System for Blood-Brain Barrier Opening in Rodents
Published on: July 16, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022