Accuracy of oxygen delivery through bubble humidifiers and nasal catheters
Morgan L Murphy1, David S Hodgson1, Nora M Bello2
1Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS, USA.
Objective:
To evaluate the equipment used for nasal insufflation of oxygen and determine its accuracy.
Study Design:
Original study.
Methods:
Oxygen delivery assemblies consisting of a flowmeter, bubble humidifier, oxygen delivery tubing and nasal insufflation catheters were assembled. Single and double catheter assemblies were made for four sizes of nasogastric feeding tubes (3.5 Fr, 5.0 Fr, 8.0 Fr and 10.0 Fr) resulting in 64 individual assemblies. A gas flow analyzer measured oxygen flow at the tip of the nasal catheter assemblies and from the pressure relief valve (PRV) of the bubble humidifiers. Statistical analyses were conducted to assess the functionality of assemblies. For functional assemblies, the accuracy of oxygen flow relative to the prescribed flow settings was determined.
Results:
Catheter size was significantly associated with the functionality of assemblies. Probability (95% confidence interval) of 3.5 Fr, 5.0 Fr and 8.0 Fr assemblies being functional was estimated at 0.53 (0.14, 0.89), 0.83 (0.36, 0.98) and 0.98 (0.76, 0.99), respectively. All 10.0 Fr assemblies were functional. Functional assemblies, in general, consistently under-delivered the prescribed flow because a large portion of set flow was diverted through the bubble humidifier PRV.
Conclusions:
Leaks through the PRV cause significant diversion of oxygen prior to it reaching the catheter tips. Smaller patients are particularly susceptible, as small catheters limit oxygen delivery creating proportionally greater leaks through the PRV.
Clinical Relevance:
It was not possible to accurately deliver oxygen because of leaks through the PRV. Targeting a specific outcome (e.g., oxyhemoglobin saturation > 94%, PaO2 80-120 mmHg; 11-16 kPa) and avoiding unnecessarily high fractions of inspired oxygen cannot be done if flow delivery cannot be accurately assured. One possible solution would be to use a bubble humidifier with a 6 psi PRV that does not leak prior to reaching the opening pressure.
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...