Investigation of Radiation Dose Estimates and Image Quality Between Commercially Available Interventional Fluoroscopy
Lukas M Trunz1, David J Eschelman1, Carin F Gonsalves1
1Department of Radiology, Thomas Jefferson University, 132 S 10th Street, Philadelphia, PA 19107.
Rationale And Objectives:
To investigate differences in radiation dose and image quality for single-plane flat-panel-detector based interventional fluoroscopy systems from two vendors using phantom study and clinical procedures.
Materials And Methods:
AlluraClarityIQ (Philips) and Artis Q (Siemens-Healthineers) interventional fluoroscopy systems were evaluated. Phantom study included comparison of system-reported air-kerma rates (AKR) for clinical protocols with simulated patient thicknesses (20-40 cm). Differences in system-reported radiation dose estimates, cumulative-air-kerma (CAK) and kerma-area-product (KAP), for different clinical procedures were investigated. Subset analysis investigated differences in CAK, KAP and other factors affecting radiation dose when the same patients underwent repeat embolization procedures performed by the same physician on the two different fluoroscopy systems. Two blinded interventional radiologists reviewed image-quality for these procedures using a five-point scale (1-5; 5-best) for five parameters.
Results:
Phantom study revealed that air-kerma rates was significantly higher for Artis Q system for 30-40cm of simulated patient thicknesses (p < 0.05). Overall data analysis from 4381 clinical cases revealed significant differences in CAK and KAP for certain procedures (p < 0.05); with significantly lower values for AlluraClarityIQ systems (median CAK lower by: 29%-58%). Subset analysis with 40 patients undergoing repeat embolization procedures on both systems revealed that median CAK and KAP were significantly lower for AlluraClarityIQ systems (p < 0.02) by 45% and 31%, respectively. Image quality scores for AlluraClarityIQ systems were significantly greater (mean difference range for five parameters: 1.3-1.6; p < 0.005).
Conclusion:
Radiation dose and image quality differences were observed between AlluraClarityIQ and Artis Q systems. AlluraClarityIQ systems showed lower radiation utilization and an increase in subjective perception of image quality.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...