Monte Carlo dose calculations of shielding disks with different material combinations in intraoperative electron
H Alhamada1, S Simon1, C Philippson2
1Nuclear Metrology, école polytechnique, université Libre de Bruxelles, Bruxelles, Belgium; Radiothérapy Department, institut Jules-Bordet, Bruxelles, Belgium.
Purpose:
Shielding disks play an important role in intraoperative electron radiation therapy, and different designs are currently used in clinical practice. This paper investigates the dosimetric impact of the shielding disk used during intraoperative electron radiation therapy (IOERT).
Materials And Methods:
This paper focuses on the study of four shielding disks types that have been used in our clinic: Aluminum (Al)/Lead (Pb), PMMA/Copper(Cu)/PMMA, Aluminum (Al)/Copper (Cu) and Aluminum (Al)/Steel with their specific thicknesses. The theoretical study was conducted with the EGSnrc Monte Carlo (MC) code. On the other hand, the measurements were carried out with gafchromic films for the four shielding disks for the same setup inside the water phantom. Finally, a comparison of the simulated and measured PDD curves was performed for the four material combinations.
Results:
MC simulation and gafchromic measurements illustrated that dose values under the four shielding disks types were close to 0, whereas the backscattering enhancement of the disks were 103% with Al/Pb shielding disk, 102% with Al/Steel shielding disk, 102% with Al/Cu shielding disk, 95% with PMMA/Cu/PMMA shielding disk. The PDDs values of the gafchromic films in front of the disks were: 107%, 105%, 104%, and 94% for the Al/Pb, Al/Steel, Al/Cu, and PMMA/Cu/PMMA disks respectively.
Conclusions:
The dose values above and under the shielding disks were acceptable for the four studied shielding types. Demonstrated it is possible to use any of them clinically, while the best shielding disk was the Al/Pb since it has minimum thickness and a small backscatter enhancement.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Biological Effects of Radiation
Diamagnetic Shielding of Nuclei: Local Diamagnetic Current
Nuclear Power
Nuclear Fuels
Nuclear fuel consists of a fissile isotope, such as uranium-235, which must be present in sufficient quantity to provide a...
Dose Size and Dosing Frequency: Determination Methods