Quality Comparison of the Manual Chest Compression and the Mechanical Chest Compression During Difficult Transport
Burak Bekgöz1, İshak Şan2, Mehmet Ergin3
1Department of Emergency Medicine, Ankara City Hospital, Ankara, Turkey.
Background:
Although there are several studies comparing the quality of manual and mechanical chest compressions, we decided to conduct this study because results of previous studies were not sufficient for us to arrive at a definite conclusion.
Objective:
In this study, our goal was to evaluate the quality of cardiopulmonary resuscitation (CPR) performed manually and by mechanical chest compression device (MCCD) when removing out-of-hospital cardiac arrest patients from their homes via stairs.
Methods:
A total of 20 paramedics participated in the study. The patient simulator manikin was moved down the stairs while each of 20 paramedics performed chest compressions, then it was moved down the stairs again 20 times while the MCCD performed chest compressions. Compression depth, compression rate, and hands-on times were recorded and the data were compared.
Results:
The median chest compression rate was 142.0 compressions/min (interquartile [25th to 75th percentile] range [IQR] 134.9-148.7 compressions/min) for the paramedics and 102.3 compressions/min for the MCCD (IQR 102.2-102.5 compressions/min) (p < 0.01). The median chest compression depth was 25.2 mm (IQR 23.2-30.9 mm) for the paramedics and 52.0 mm for the MCCD (IQR 51.4-52.6 mm) (p < 0.001). The rate of hands-on time for chest compressions performed by the paramedic participants was 92.0% (IQR 86.5-100%). Hands-on rate of the MCCD was 100% (p = 0.09).
Conclusions:
In our study, while carrying the patient simulator manikin to the lower floor, it was found that the MCCD achieved high-quality CPR targets recommended by resuscitation guidelines in terms of compression rate, depth, and hands-on-time.
Related Concept Videos
Cardiopulmonary Resuscitation I: Adult
Cardiopulmonary Resuscitation II: ACLS Airway Management
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Flail Chest-II
Assessment:
1. Clinical Evaluation:
History:
Assessment of the Cardiovascular System III: Palpation
Jugular Venous Pressure (JVP) Measurement
Position the patient at a thirty- to forty-five-degree angle or in a semi-fowler's position. Look for the highest point of pulsation in the internal jugular vein and measure the vertical distance to the angle of Loius or sternal angle. A normal JVP is 3-4 cm above...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...