CORRELATION OF EYE LENS DOSES AND PERSONAL DOSE EQUIVALENT MEASURED ON THE ARM OF INTERVENTIONAL CARDIOLOGISTS FOR A
Joanna Domienik-Andrzejewska1, Marcin Brodecki1, Marek Zmyślony1
1Radiation Protection Department, Nofer Institute of Occupational Medicine, 8 St. Teresa Str, Łódź 91-348, Poland.
Abstract:
Coefficients converting the readings of the whole body dosemeter worn on the left arm to eye lens doses were determined by analysing the correlations between Hp(10) and Hp(3) values. Doses were measured on a phantom for specific C-arm projections typically used during CA/PCI procedures. In order to estimate the cumulative eye lens doses, conversion coefficients were then applied to the dose records of interventional cardiologists collected in the database of dosimetry service between the years 1995 and 2009. The Hp(10) to Hp(3) conversion coefficients are 0.29 (CV = 34%) and 0.17 (CV = 42%) for left and right eye lens, respectively. However, they can vary from one laboratory to another depending on working technique. From among 61 interventional cardiologists, none exceeded the threshold dose of 0.5 Gy for eye lens opacities. However, 44% of interventional cardiologists were likely to exceed the annual limit of 20 mSv for the most exposed eye at least once in the analysed time period.