ISO/TS 20914:2019 - a critical commentary
Ian Farrance1, Robert Frenkel2,3, Tony Badrick4
1Discipline of Laboratory Medicine, School of Health and Biomedical Sciences, RMIT University, Victoria, Australia.
The 2019 ISO/TS 20914 offers guidance for measurement uncertainty in medical labs but contains statistical inaccuracies. Key concerns include misuse of coefficient of variation and flawed calculations for eGFR and INR.
Area of Science:
- Clinical Laboratory Science
- Metrology
- Biostatistics
Background:
- ISO/TS 20914, published in July 2019, provides practical guidance for estimating measurement uncertainty in medical laboratories.
- The document claims to be based on the Evaluation of measurement data - Guide to the expression of uncertainty in measurement (GUM), JCGM 100:2008.
- Despite its detailed examples, the guide presents several potentially misleading or incorrect statistical procedures.
Purpose of the Study:
- To critically evaluate the statistical methodologies and statements within ISO/TS 2019.
- To identify and highlight major concerns regarding the estimation of measurement uncertainty in medical laboratories as presented in the ISO guide.
- To provide a detailed commentary on specific problematic areas within the ISO/TS 20914 document.
Main Methods:
- Comparative analysis of ISO/TS 2019 with established metrological principles (GUM).
- Identification and detailed examination of specific statistical procedures and statements within the ISO guide.
- Case study analysis of calculations for estimated glomerular filtration rate (eGFR) and international normalized ratio (INR).
Main Results:
- Concerns identified regarding the use and potential misuse of coefficient of variation.
- Oversimplification and potential misleading nature of pooled variance and measurement uncertainty across changing conditions.
- Incomplete inclusion of all known uncertainties in estimated glomerular filtration rate (eGFR) calculations.
- Incorrect international normalized ratio (INR) calculation identified.
- Problematic treatment of bias uncertainty and incomplete rules for evaluating combined uncertainty in functional relationships.
Conclusions:
- ISO/TS 2019, while offering practical guidance, contains significant statistical inaccuracies and potentially misleading information.
- Clarification and correction are needed for several statistical procedures, including coefficient of variation, pooled variance, bias uncertainty, and combined uncertainty calculations.
- The accuracy of eGFR and INR calculations as presented in the guide is questionable, requiring further review and revision to ensure reliable measurement uncertainty estimation in medical laboratories.
More Related Videos
10:24Applicability Analysis of Assessment Methods for Morphological Parameters of Corroded Steel Bars
Published on: November 1, 2018
05:47Simulation of a Scaled Assembly Process with Collaboration of a Robotic Arm and Monitoring through a Vision System for Quality Control
Published on: August 29, 2025
Related Concept Videos
Legal Guidelines for Documentation
Isothermal Processes
An ideal gas can also undergo isothermal expansion or compression.
For example, consider 1 mole of an ideal gas inside an isolated cylinder at initial volume V...
Methods of Documentation I: Source-Oriented Records
In an SOR, each discipline involved in patient care maintains a separate medical record section. This record-keeping method enables easy tracking of patient progress and ensures healthcare staff have access to up-to-date information.
Key Attributes include the following:
Isochoric and Isobaric Processes
Suppose 1000 g of water is heated from 40...
SI Units: 2019 Redefinition
A standard set of units has been defined...
Methods of Documentation V: CBE
In CBE, healthcare professionals establish predefined standards of practice that define what constitutes...