Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
16.1K
Intensity-modulated radiation therapy for multiple targets with tomotherapy using multiple sets of static ports from
Yoshihiko Manabe1, Yuta Shibamoto1, Taro Murai1
1Department of Radiology, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan.
Journal of Applied Clinical Medical Physics
|April 7, 2020
Summary
The novel pseudo-dynamic jaw displacement (pDJDC) technique significantly reduces tomotherapy treatment time for multiple targets by 48% compared to helical tomotherapy (HDJ). While dose distribution is comparable, the pDJDC technique shows a slightly reduced conformity index.
Area of Science:
- Radiation Oncology
- Medical Physics
- Cancer Treatment
Background:
- A new pseudo-dynamic jaw displacement (pDJDC) technique was developed for rapid treatment of multiple, spatially separated targets using static-port tomotherapy.
- This technique was compared against helical tomotherapy using the dynamic-jaw mode (HDJ) for treating multiple targets.
Purpose of the Study:
- To evaluate the efficacy and safety of the pDJDC technique compared to the HDJ technique for treating multiple targets.
- To assess treatment time, dose distribution, conformity, and uniformity.
Main Methods:
- Seven patients with multiple targets were treated with both pDJDC and HDJ techniques.
- Plans were compared based on conformity index, uniformity index (D5%/D95%), lung dose distribution (mean lung dose, V5Gy, V20Gy), and treatment duration.
Main Results:
- The pDJDC technique reduced treatment time by 48% (median 462s vs. 884s, P=0.031).
- Conformity index was slightly worse for pDJDC (3.0 vs. 2.4, P=0.031), but uniformity index was comparable (1.048 vs. 1.057, P=0.10).
- Lung dose parameters were similar between the two techniques.
Conclusions:
- The pDJDC technique offers a significant reduction in treatment time for multiple targets.
- While dose distribution is comparable, the pDJDC technique results in a slightly lower conformity index.
- pDJDC is a promising technique for efficient multi-target radiotherapy.