Patient size matters: Effect of tube current modulation on size-specific dose estimates (SSDE) and image quality in
Izabella Barreto1, Nupur Verma1, Nathan Quails1
1Department of Radiology, College of Medicine, University of Florida, Gainesville, FL, USA.
Purpose:
We compare the effect of tube current modulation (TCM) and fixed tube current (FTC) on size-specific dose estimates (SSDE) and image quality in lung cancer screening with low-dose CT (LDCT) for patients of all sizes.
Methods:
Initially, 107 lung screening examinations were performed using FTC, which satisfied the Centers for Medicare & Medicaid Services' volumetric CT dose index (CTDIvol ) limit of 3.0 mGy for standard-sized patients. Following protocol modification, 287 examinations were performed using TCM. Patient size and examination parameters were collected and water-equivalent diameter (Dw ) and SSDE were determined for each patient. Regression models were used to correlate CTDIvol and SSDE with Dw . Objective and subjective image quality were measured in 20 patients who had consecutive annual screenings with both FTC and TCM.
Results:
CTDIvol was 2.3 mGy for all FTC scans and increased exponentially with Dw (range = 0.96-4.50 mGy, R2 = 0.73) for TCM scans. As patient Dw increased, SSDE decreased for FTC examinations (R2 = 1) and increased for TCM examinations (R2 = 0.54). Image quality measurements were superior with FTC for smaller sized patients and with TCM for larger sized patients (R2 > 0.5, P < 0.005). Radiologist graded all images acceptable for diagnostic evaluation of lung cancer screening.
Conclusion:
Although FTC protocol offered a consistently low CTDIvol for all patients, it yielded unnecessarily high SSDE for small patients and increased image noise for large patients. Lung cancer screening with LDCT using TCM produces radiation doses that are appropriately reduced for small patients and increased for large patients with diagnostic image quality for all patients.
More Related Videos
07:56Evaluating Regional Pulmonary Deposition using Patient-Specific 3D Printed Lung Models
Published on: November 11, 2020
11:31Using Micro-computed Tomography for the Assessment of Tumor Development and Follow-up of Response to Treatment in a Mouse Model of Lung Cancer
Published on: May 20, 2016
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT