Background:
Invasive coronary angiography (ICA) still causes a significant amount of radiation exposure for patients and operators. In February 2017, the Azurion system was introduced, a new-generation fluoroscopy image acquisition and processing system. Radiation exposure in patients undergoing ICA was assessed comparing the novel Azurion 7 F12 angiography system to its predecessor Allura Xper in a randomized manner.
Methods:
Radiation exposure was prospectively analyzed in 238 patients undergoing diagnostic ICA. Patients were randomly assigned to the novel Azurion system (119 patients) or its predecessor Allura Xper system (119 patients). In each patient, 8 predefined standard projections (5 left coronary artery, 3 right coronary artery) were performed. Image quality was quantified by grading of the images on the basis of a 5-point grading system.
Results:
Radiation dose area product was significantly lower in the Azurion group 109 (interquartile range [IQR 75-176] cGy cm) compared with the Allura Xper group 208 [IQR 134-301] cGy cm (P<0.001). Body mass index (26.6 [IQR 23.9-29.7] kg/m vs. 26.2 [IQR 24.2-29.4] kg/m; P=0.607), body surface area (1.96 [IQR 1.81-2.11] m vs. 1.90 [IQR 1.77-20.4] m; P=0.092), and procedure duration (1.5 [IQR 1.2-2.3] min vs. 1.6 [IQR 1.2-2.5] min; P=0.419) were similar in both groups. Images from the Azurion system were at least of equal quality compared with Allura Xper (image quality grade 4.82±0.45 vs. 4.75±0.52, P=0.43).
Conclusion:
Use of the novel Azurion 7 F12 angiography system resulted in a significant reduction of dose area product in patients undergoing diagnostic ICA by 56%.
More Related Videos
10:46Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
13:48Reduction of Radiation Exposure during Endovascular Treatment of Peripheral Arterial Disease Combining Fiber Optic RealShape Technology and Intravascular Ultrasound
Published on: April 21, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies VII: Vascular Imaging
Radiological Investigation I: X-ray and CT
Cardiac Catheterization I: Pre-Procedure Overview