Related Experiment Video
Updated: Dec 24, 2025

Technique of Porcine Liver Procurement and Orthotopic Transplantation using an Active Porto-Caval Shunt
Published on: May 7, 2015
MIEO: a micro-invasive endoscopic operation port system for transluminal interventions-an acute and survival porcine
D Wilhelm1,2, T Vogel3,4, A Jell3,4
1Technische Universität München, Fakultät für Medizin, Klinik Und Poliklinik für Chirurgie, Klinikum Rechts Der Isar, Ismaningerstr. 22, München, 81675, Germany. dirk.wilhelm@tum.de.
Background:
A reliable and sterile access through the intestinal wall to ease flexible endoscopic transluminal interventions is still appealing but lacks a suitable port system.
Methods:
In a granted industry cooperation, we developed the MIEO-Port, a flexible three components overtube system that provides a temporary hermetic sealing of the intestinal wall to allow endoscopic disinfection and manipulation to gain access to the abdominal cavity. The port features an innovative head part which allows for coupling the port to the intestinal wall by vacuum suction and for controlled jetting the isolated intestinal surface with a disinfectant. The device was tested in vivo in 6 pigs for acute and long-term usability. All animal tests were approved by the local ethics committee.
Results:
In the acute experiment, the port system supported sealed endoscopic mucosa resection and transluminal cholecystectomy. In the survival study on 5 animals, the MIEO-Port proved its reliability after transcolonic peritoneoscopy. In one animal, a port dislocation occurred after extensive retroperitoneal preparation, one animal revealed bacterial contamination at necropsy; however, all animals showed a favourable course over ten days and offered no signs of peritonitis or abscedation during post-mortem examination.
Discussion:
To the best of our knowledge, the MIEO-Port system is the first device to provide a reliable and sterile flexible access to the peritoneal cavity that can be used throughout the entire gastrointestinal tract regardless of the access route and which combines hermetic sealing with local sterilization. Further studies are warranted.