An electromagnetic separation system for the enrichment of 39Ar
Z H Jia1, Amin L Tong2, L T Sun1
1Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China.
Abstract:
An isotope enrichment system for 39Ar has been developed at the Institute of Modern Physics, which is designed to increase the abundance of 39Ar in the incident sample gas. With intense Ar+ beams produced by a 2.45 GHz electron cyclotron resonance ion source and a high mass resolution spectrometer system, Ar isotopes are evidently separated on the target plane and selectively collected by an Al target. The separated Ar isotopes have been identified on the target plane, which is consistent with the simulations. According to the recent cross-checked results with atom trap trace analysis, a high enrichment factor of 39Ar has been successfully achieved. This paper will present the design and test results of this system.