Three-dimensional Optical-resolution Photoacoustic Microscopy
Published on: May 3, 2011
Improved Depth-of-Field Photoacoustic Microscopy with a Multifocal Point Transducer for Biomedical Imaging
Thanh Phuoc Nguyen1, Van Tu Nguyen2, Sudip Mondal3
1Department of Mechatronics, Cao Thang Technical College, Ho Chi Minh City 700000, Vietnam.
Abstract:
In this study, a photoacoustic microscopy (PAM) system based on a multifocal point (MFP) transducer was fabricated to produce a large depth-of-field tissue image. The customized MFP transducer has seven focal points, distributed along with the transducer's axis, fabricated by separate spherically-focused surfaces. These surfaces generate distinct focal zones that are overlapped to extend the depth-of-field. This design allows extending the focal zone of 10 mm for the 11 MHz MFP transducer, which is a great improvement over the 0.48 mm focal zone of the 11 MHz single focal point (SFP) transducer. The PAM image penetration depths of a chicken-hemoglobin phantom using SFP and MFP transducers were measured as 5 mm and 8 mm, respectively. The significant increase in the PAM image-based penetration depth of the chicken-hemoglobin phantom was a result of using the customized MFP transducer.