Related Experiment Video
Updated: Dec 24, 2025

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Diagnosis of sleep apnea without sensors on the patient's face
AbdelKebir Sabil1, Caroline Marien2, Marc LeVaillant3
1Cloud Sleep Lab, Paris, France.
Study Objectives:
Thermistors, nasal cannulas, and respiratory inductance plethysmography (RIP) are the recommended reference sensors of the American Academy of Sleep Medicine (AASM) for the detection and characterization of apneas and hypopneas; however, these sensors are not well tolerated by patients and have poor scorability. We evaluated the performance of an alternative method using a combination of tracheal sounds (TSs) and RIP signals.
Methods:
Consecutive recordings of 70 adult patients from the Pays de la Loire Sleep Cohort were manually scored in random order using the AASM standard signals and the combination TS and RIP signals, without respiratory sensors placed on the patient's face. The TS-RIP scoring used the TS and RIP-flow signals for detection of apneas and hypopneas, respectively, and the suprasternal pressure and RIP belt signals for the characterization of apneas.
Results:
Sensitivity and specificity of the TS-RIP combination were 96.21% and 91.34% for apnea detection and 89.94% and 93.25% for detecting hypopneas, respectively, with a kappa coefficient of 0.87. For the characterization of apneas, sensitivity and specificity were 98.67% and 96.17% for obstructive apneas, 92.66% and 99.36% for mixed apneas, and 96.14% and 98.89% for central apneas, respectively, with a kappa coefficient of 0.94. The TS-RIP scoring revealed a high agreement for classifying obstructive sleep apnea into severity classes (none, mild, moderate, and severe obstructive sleep apnea) with a Cohen's kappa coefficient of 0.96.
Conclusions:
Compared with the AASM reference sensors, the TS-RIP combination allows reliable noninvasive detection and characterization of respiratory events with a high degree of sensitivity and specificity. TS-RIP combination could be used for diagnosis of obstructive sleep apnea in adults, either as an alternative to the AASM sensors or in combination with the recommended AASM sensors.
Related Concept Videos
Sleep Apnea
The condition is more prevalent among...
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Cardiopulmonary Resuscitation II: ACLS Airway Management
Physical Assessment of the Respiratory Tract I: Health History
Subjective Data
Subjective data provides vital information about the patient's health history and symptoms. This data is typically collected through interviews in which patients describe their experiences, symptoms, and concerns.
Health history and...