Related Experiment Video
Updated: Dec 24, 2025

Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
Investigating the estimation of cardiac time intervals using gyrocardiography
Parastoo Dehkordi1, Kouhyar Tavakolian2, Mojtaba Jafari Tadi3
1Electrical and Computer Engineering Department, University of British Columbia, Vancouver, British Columbia, Canada.
Objective:
Assessment of cardiac time intervals (CTIs) is essential for monitoring cardiac performance. Recently, gyrocardiography (GCG) has been introduced as a non-invasive technology for cardiac monitoring. GCG measures the chest's angular precordial vibrations caused by myocardium wall motion using a gyroscope sensor attached to the sternum. In this study, we investigated the accuracy and reproducibility of estimating CTIs from the GCG recordings of 50 adults.
Approach:
We proposed five fiducial points for the GCG waveforms associated with the opening and closure of aortic and mitral valves. Two annotators annotated the suggested points on each cardiac cycle. The points were compared to the corresponding opening and closing of cardiac valves delineated on Tissue Doppler imaging (TDI) recordings. The fiducial points were annotated on seismocardiography (SCG) and impedance cardiography (ICG) signals recorded simultaneously.
Main Results:
For estimating the timing of mitral valve closure, aortic valve opening, aortic valve closure, and mitral valve opening, 40%, 67%, 75%, and 70% of GCG annotations fell in the corresponding echocardiography ranges, respectively. The results showed moderate-to-excellent (r = 0.4-0.92; p-value < 0.01) correlation between the measured and the reference CTls. A myocardial performance index (Tei index) adapted using joint GCG and SCG resulted in a moderate correlation (r = 0.4; p-value < 0.001).
Significance:
The findings showed that the CTIs can be easily measured using GCG. Also, we found that using SCG and GCG recordings together could provide an opportunity to estimate CTIs more accurately, and make it possible to calculate the Tei index as an indicator of myocardial performance.
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Dysrhythmias V: Evaluating Dysrhythmias
Cardiac Cycle
During the cardiac cycle, blood flow through the heart is regulated entirely by changing pressure gradients. This sequence of events begins with the heart in a state of total relaxation, known as mid-to-late diastole, during which blood passively flows from...