Effect of acquisition depth and precompression from probe and couplant on shear wave elastography in soft tissue: an
Xiuming Wang1, Yue Hu1, Jia'an Zhu1
1Department of Ultrasound, Peking University People's Hospital, 11 Xizhimen South Street, Beijing 100044, China.
Background:
Shear wave elastography is a promising method to diagnose early musculoskeletal lesions. We aimed to explore the feasible depth and acceptable precompression applied by probe and couplant for soft tissues in the present system.
Methods:
Ex-vivo muscles were evaluated at depths of 0.5-6 cm by 3 operators, using 1-5 mm couplant thickness and 0-3.0 kPa probe pressure. We compared the shear wave speed (SWS) and used intraclass correlation coefficients to assess reproducibility. In vivo skin and subcutaneous superficial fascia from volunteers were tested at depths ranging from 0.1-0.5 cm with 1-20 mm couplant thickness.
Results:
The SWS of ex-vivo muscles varied and increased with depth, and could not be acquired at 6 cm because the shear wave failed to be detected. Furthermore, while the SWS of ex-vivo muscles were not affected by the couplant thickness, it was affected by probe pressure. Most cases demonstrated a satisfactory agreement degree of the intraoperator reproducibility (ICC, 0.81-0.95) and a substantial interoperator reproducibility (ICC >0.60). Inter- and intra-operator reproducibility was better at a depth of 0.5-4 cm than at 5 cm. In the in vivo study, when tissues within a 0.2 cm depth were evaluated, the SWS that was acquired using a couplant thickness of >10 mm was different from that acquired using other thicknesses.
Conclusions:
The SWS acquired at a depth of ≤3 cm with a suitable amount of couplant is recommended.
More Related Videos
07:57Viscoelastic Characterization of Soft Tissue-Mimicking Gelatin Phantoms using Indentation and Magnetic Resonance Elastography
Published on: May 10, 2022
09:02Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):