Single-channel oximetry monitor versus in-lab polysomnography oximetry analysis: does it make a difference?
Joachim A Behar1, Niclas Palmius2, Sroussi Zacharie1
1Faculty of Biomedical Engineering, Technion, Israel Institute of Technology, Haifa, Israel.
Objective:
Portable oximetry has been shown to be a promising candidate for large-scale obstructive sleep apnea screening. In polysomnography (PSG), the gold standard OSA diagnosis test, the oxygen desaturation index (ODI) is usually computed from desaturation events occurring during sleep periods only, i.e. overnight desaturations occurring during or overlapping with a wake state are excluded. However, for unattended home oximetry, all desaturations are taken into account since no reference electroencephalogram is available for sleep staging. We aim to evaluate the hypothesis that the predictive power of oximetry for OSA screening is not impaired when reference sleep stages are not available.
Approach:
We used a PSG clinical database of 887 individuals from a representative São Paulo (Brazil) population sample. Using features derived from the oxygen saturation time series and demographic information, OxyDOSA, a published machine learning model, was trained to distinguish between non-OSA and OSA individuals using the ODI computed while including versus excluding overnight desaturations overlapping with a wake period, thus mimicking portable and PSG oximetry analyses, respectively.
Main Results:
When excluding wake desaturations, the OxyDOSA model had an AUROC = 94.9 ± 1.6, Se = 85.9 ± 2.8, Sp = 90.1 ± 2.6 and F1 = 86.4 ± 2.7. When considering wake desaturations, the OxyDOSA model had an AUROC = 94.4 ± 1.6, Se = 88.0 ± 2.0, Sp = 87.7 ± 2.9 and F1 = 86.2 ± 2.4. Non-inferiority was demonstrated (p = 0.049) at a tolerance level of 3%. In addition, analysis of the desaturations excluded by PSG oximetry analysis suggests that up to 21% of the total number of desaturations might actually be related to apneas or hypopneas.
Significance:
This analysis of a large representative population sample provided strong evidence that the predictive power of oximetry for OSA screening using the OxyDOSA model is not impaired when reference sleep stages are not available. This finding motivates the usage of portable oximetry for OSA screening.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...