Related Concept Videos

Errors occurring during blood pressure monitoring01:25

Errors occurring during blood pressure monitoring

1.2K
Blood pressure monitoring is a crucial clinical procedure in diagnosing and managing various cardiovascular conditions. Despite its significance, the accuracy of blood pressure measurements can be compromised by multiple factors, potentially leading to either falsely high or low readings. These inaccuracies are critical as they can significantly impact patient care. So, it is vital to understand these challenges deeply and adopt strategic approaches to minimize errors.
Several factors...
1.2K
Measurement of Blood Pressure01:17

Measurement of Blood Pressure

2.4K
Assessing blood pressure is a standard procedure executed in virtually all medical environments. The method utilized today was established over a hundred years ago by an innovative Russian doctor, Dr. Nikolai Korotkoff. The soft ticking noise, known as Korotkoff sounds, heard while taking blood pressure readings results from turbulent blood flow within the vessels. The apparatus required for this procedure includes a sphygmomanometer, a blood pressure cuff attached to a gauge, and a...
2.4K
Assessment of blood pressure in brachial artery(two-step method)01:23

Assessment of blood pressure in brachial artery(two-step method)

1.4K
Measuring blood pressure is a fundamental skill in healthcare that aids in diagnosing and monitoring hypertension and other cardiovascular conditions. An aneroid sphygmomanometer, commonly used in clinical settings, offers a manual and precise method for blood pressure measurement. The technique for using this instrument involves specific steps that must be carefully executed to ensure accuracy. The following detailed description outlines a two-step technique for assessing blood pressure using...
1.4K
Equipments Used To Measure Blood Pressure01:30

Equipments Used To Measure Blood Pressure

2.9K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.9K
Sites for measruring blood pressure01:21

Sites for measruring blood pressure

3.1K
Blood pressure measurement is a fundamental clinical procedure, providing crucial data for assessing cardiovascular health. Among the various sites for this measurement, the brachial and popliteal arteries are predominantly utilized due to their accessibility and the reliability of their readings. This lesson delves into the anatomical significance, methodology, and considerations of measuring blood pressure at these locations.
The Brachial Artery: Primary Site for Blood Pressure Measurement
3.1K
Special considerations while measuring blood pressure01:28

Special considerations while measuring blood pressure

1.0K
When assessing blood pressure (BP), healthcare professionals must consider various factors and potential unexpected outcomes to ensure accurate readings and provide proper patient care. Adhering to these guidelines is essential to achieving the most reliable results.
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
1.0K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Evaluating neural encoding of prosody-related F0 changes in emotional speech using the speech FFR in normal-hearing adults.

Scientific reports·2026
Same author

Pathways to Wellbeing: Reconceptualizing Resilience to Foreground Marginalized Teachers' Agentic Resistance.

Behavioral sciences (Basel, Switzerland)·2025
Same author

Revisiting the relationship between pulse transit time and arterial stiffness.

Pflugers Archiv : European journal of physiology·2025
Same author

A method for pitch tracking in the frequency following response using harmonic amplitude summation filterbank.

JASA express letters·2025
Same author

Investigating Performance of an Embedded Machine Learning Solution for Classifying Postural Behaviors.

Sensors (Basel, Switzerland)·2025
Same author

Imbalanced feature generation based on bootstrap power spectral curve for estimating respiratory rate.

Scientific reports·2025

Related Experiment Video

Updated: Dec 24, 2025

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
14:28

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver

Published on: June 27, 2025

819

Uncertainty in Blood Pressure Measurement Estimated Using Ensemble-Based Recursive Methodology.

Soojeong Lee1, Hilmi R Dajani2, Sreeraman Rajan3

  • 1Department of Computer Engineering, Sejong University, 209 Neungdong-ro, Gwangjin-gu, Seoul 05006, Korea.

Sensors (Basel, Switzerland)
|April 12, 2020
PubMed
Summary

This study introduces a new method to measure blood pressure uncertainty using automated oscillometric monitors. The technique enhances accuracy for systolic and diastolic blood pressure readings.

Keywords:
confidence intervaldeep neural networkensemble methodoscillometry blood pressure measurementuncertainty

More Related Videos

Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
11:26

Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression

Published on: December 10, 2014

12.7K
Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
08:54

Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method

Published on: May 15, 2009

53.0K

Related Experiment Videos

Last Updated: Dec 24, 2025

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver
14:28

Software for Analysis of Heart Rate and Blood Pressure Time-series Data from the Valsalva Maneuver

Published on: June 27, 2025

819
Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
11:26

Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression

Published on: December 10, 2014

12.7K
Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
08:54

Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method

Published on: May 15, 2009

53.0K

Area of Science:

  • Biomedical Engineering
  • Medical Devices
  • Signal Processing

Background:

  • Automated oscillometric blood pressure monitors are widely used but lack uncertainty quantification.
  • Current devices provide single readings without indicating measurement reliability.
  • Accurate blood pressure monitoring is crucial for patient health management.

Purpose of the Study:

  • To develop a novel technique for quantifying measurement uncertainty in oscillometric blood pressure devices.
  • To improve the reliability and interpretability of automated blood pressure measurements.
  • To provide clinicians and patients with a more comprehensive understanding of blood pressure readings.

Main Methods:

  • An ensemble-based recursive methodology is proposed.
  • A three-stage approach includes pre-learning with bagging and parameter fine-tuning.
  • The method estimates both blood pressure and its associated uncertainty.

Main Results:

  • The proposed technique enables the measurement of uncertainty for systolic and diastolic blood pressure.
  • The ensemble-based recursive methodology refines parameter estimation for improved accuracy.
  • This approach addresses the limitation of single-point measurements in current devices.

Conclusions:

  • The developed methodology offers a significant advancement in automated blood pressure monitoring.
  • Quantifying uncertainty enhances the clinical utility of oscillometric devices.
  • This technique can lead to more informed diagnostic and treatment decisions.