Doppler Optical Coherence Tomography of Retinal Circulation
Published on: September 18, 2012
Multiple scan averaging to yield accurate quantitative analysis of optical coherence tomography angiograms
Hafi M Khan1, Alex Gentle1, James A Armitage1
1School of Medicine, Faculty of Health, Deakin University, Deakin, Australia.
Abstract:
Optical coherence tomography angiography (OCTA) is widely used in ophthalmic practice. Most OCTA studies based their findings on a single OCTA measurement. We conducted an observational study of 82 eyes from 82 healthy subjects to compare variations of OCTA parameters among five successive measurements. A 3 × 3 mm Early Treatment of Diabetic Retinopathy Study grid centred at fovea was used. An average from five successive OCTA measurements (both perfusion density and vessel density) was calculated to be used as the reference standard. There was no significant difference in perfusion and vessel densities among five successive OCTA measurements, and from different levels of averaging. Perfusion density was close to the reference standard when average from three measurements was used (discrepancy within 1.5%) as compared with using just one measurement (discrepancy from 3.2% to 4.5%). Vessel density was also close to reference standard when average from three measurements was used (within 0.8 mm-1) as compared with using just one measurement (2 mm-1). Software feature that allows OCTA devices to average quantitative parameters for analysis will be useful.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies VII: Vascular Imaging
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT