Optimising image quality and radiation dose for neonatal incubator imaging
J Tugwell-Allsup1, R W Morris1, R Hibbs2
1Besti Cadwaladr University Health Board, Ysbyty Gwynedd, Penrhosgarnedd Road, LL57 2PW, Bangor, Gwynedd, UK.
Radiography (London, England : 1995)
|April 14, 2020
Summary
Optimizing neonatal chest radiography in incubators is crucial. Direct imaging improves image quality and reduces radiation dose compared to using incubator trays, which requires parameter adjustments.
Area of Science:
- Medical Imaging
- Radiography
- Neonatal Care
Background:
- Neonatal radiography often occurs within incubators.
- Limited evidence exists on optimal imaging methods and parameters for neonates in incubators.
- Standardizing and optimizing these examinations is essential.
Purpose of the Study:
- To standardize and optimize neonatal chest radiography within incubators.
- To evaluate the impact of different acquisition parameters on image quality and radiation dose.
- To provide evidence-based recommendations for clinical practice.
Main Methods:
- A neonatal anthropomorphic phantom was imaged in two incubators using a DR system.
- Varied exposure factors, source-to-image distance (SID), and image receptor placement (direct vs. tray).
- Assessed image quality using visual grading analysis (VGA) and contrast-to-noise ratio (CNR); calculated effective dose via Monte Carlo simulation.
Main Results:
- Using the incubator tray significantly reduced VGA and CNR (p < 0.05) while increasing effective dose (p < 0.05).
- Optimal parameters: direct image receptor, 0.5 mAs, 60 kV, 100 cm SID.
- If tray is necessary, adapt parameters to 1 mAs at maximum achievable SID; highest effective dose occurred with tray and 100 cm SID.
Conclusions:
- Incubator trays negatively impact image quality and increase radiation dose during neonatal imaging.
- Direct imaging is optimal; if trays are used, mAs and SID require adjustment.
- Findings inform practice for standardized and optimized neonatal incubator chest imaging.